Chapter 9: Service Maintenance and Repair
9-8
Oximeter Operation and Service Manual
Automatic shut off.
Turn on oximeter and
place finger in sensor.
Remove finger from
sensor.
Remove sensor from
oximeter. Turn on
oximeter.
Oximeter should turn
off 2 minutes (± 5 sec-
onds) after finger is
removed from sensor.
Oximeter should turn
off in 2 minutes (± 5
seconds).
Current draw through
J9-1 (total battery cur-
rent).
Oximeter on, sensor
attached to oximeter, no
finger in the sensor,
with battery voltage at
3.5 VDC, ±0.2 VDC.
110 mA, ± 15 mA