Chapter 5: Operator’s Maintenance
5-2
Oximeter Operation and Service Manual
Disposable Sensors
Disposable sensors are for single-patient use only. Disposable sensors can
be repositioned on the same patient as long as the correct SpO
2
and pulse
rate values are shown after repositioning. Disposable sensors cannot be
cleaned or repaired; discard them if they become soiled or no longer work.
Cleaning the Oximeter’s Surfaces
Clean the oximeter’s surfaces with a soft cloth moistened in a mild soap
solution. If disinfection is required, wipe the oximeter’s surfaces with a
soft cloth moistened in isopropyl alcohol. Do not allow any liquid to enter
any of the oximeter’s openings.
Long Term Storage
Remove the batteries from the oximeter before long term storage, or if the
oximeter won't be used for 6 months or more. This protects the oximeter
from damage due to batteries leaking acid.
Store the oximeter in its original shipping carton and packing materials to
help protect the oximeter from damage during storage.
CAUTION! Do not autoclave, ethylene oxide sterilize, or
immerse the oximeter in liquid.