Chapter 9: Service Maintenance and Repair
Oximeter Operation and Service Manual
9-9
Waveform Test Points
The following oscilloscope screens show typical waveforms from the sen-
sor detector’s amplifier and data acquisition circuits. Unless otherwise
noted:
•
The waveforms are measured with respect to ground at J9-2.
•
The oscilloscope is triggered on T2, falling edge.
•
The waveforms are measured with the finger sensor attached to the
oximeter, but no finger inserted in the sensor.
Connect CH1 probe to T10, T21, T22, T24. Verify the signal shown in the
respective figures.
Measurement at T2
2V/DIV
0.4mS/Div
0V