S E C T I O N 1 0
•
M a i n t e n a n c e
HOTLINE
®
3 Blood and Fluid Warmer | Operator’s Manual
31
General Inspection
Check the condition of the HOTLINE
3 Warmer with a visual
inspection before using. Remove from service any HOTLINE
Warmer
that shows physical damage.
Maintenance Performed Every 30 Days
Disinfect the Reservoir and Change Recirculating Solution for
Distilled Water and 35% Isopropyl Alcohol Solution
Refer to
Disinfect the Reservoir and Change the Recirculating Solution
procedure in this section.
Maintenance Performed Every 12 Months
Disinfect the Reservoir and Change Recirculating Solution for
0.3% Hydrogen Peroxide Solution
Refer to
Disinfect the Reservoir and Change the Recirculating Solution
procedure in this section.
Disinfect the Reservoir and Change the Recirculating
Solution
Disinfect the reservoir and change the recirculating solution every 30 days
or every 12 months based on the recirculating solution used for the
HOTLINE
3
Warmer. Refer to the following table for the maintenance
schedule
.
Recirculating Solution
Preparation
Maintenance
0.3% Hydrogen Peroxide
Solution
Mix 140 ml of 3% hydrogen
peroxide with 1,260 ml of
distilled water.
Replace solution and
disinfect reservoir every
12 months.
Distilled
Water
Use distilled water.
Replace solution and
disinfect reservoir every
30 days.
35% Isopropyl Alcohol
Solution
Mix 700 ml of 70% isopropyl
alcohol with 700 ml of distilled
water.
Replace solution and
disinfect reservoir every
30 days.
Note: Use distilled water only, not tap water. Failure to do so may cause build-up of
mineral deposits in the recirculating solution path, which may impair heater
performance.