- 20 -
Language /
语言
Switch the system Language
:
Chinese /English
Info
Instrument Info
Info
display
instrument
type
and the connected probe
type.
Notes:
Push the buttons firmly, but not too hard, to avoid damaging the
circuit boards.
When the instrument is used for pump control, please avoid to
directly connect the relays to pump power supply circuit.
When there is RS485 serial communication, the relay No.6 is
invalid.
The remote control can also be used to configure the parameters.
There are four keys SET
、
MOV
、
INC
、
SEL. MOV
、
INC and SEL
three keys have same functions as they are in the host panel. SET
key is the same function of MOV SEL press together in short .
5. Calibration
The instrument should be calibrate indoor before installation to ensure the
normal performance.
5.1 General consideration
Let the probe be perpendicular to a wall, and make sure the
measuring distance is larger than the blacking distance, and no
barriers within the beam angle zone.
Wiring and connecting the instrument correctly according to the
guide .
Turn on the power and after a few seconds the instrument will
enter the running mode. Check the level value, sensor state,and
press SEL key to check liquid level value , distance value and
temperature value.
Move the probe slowly , the level value and distance value should
change slowly accordingly.