background image

Can be changed

10

1000

At the time of shipment

1

5

0

1050

100

10% of rated

flow range

100% of rated

flow range

Maximum

display range

Flow [L/min]

1000 L/min type

Analog output [V]

 

Analog output free range function

Allows the flow that generates an output of 5 V or 20 mA to be changed.
The value can be changed 10% of maximum rated flow to maximum 
display value.

 

Error display function

When an error or abnormality arises, the location and contents are displayed.

If the failure cannot be solved after the above instructions are performed, please contact SMC for investigation.

Display

Description

Contents

Action

OUT1 over current 

error

Load current of 80 mA or more is applied 

to the switch output (OUT1).

Eliminate the cause of the over current by 
turning off the power supply and then turn 
on it again.

OUT2 over current 

error

Load current of 80 mA or more is applied 

to the switch output (OUT2).

Instantaneous flow 

error

The flow rate exceeds the upper limit of 

indicated flow rate range.

Decrease the flow rate.

Reverse flow error

There  is  a  reverse  flow  equivalent  to 

–5% or more.

Turn the flow to correct direction.

PFMC7501
PFMC7102
PFMC7202

Accumulated flow

error

The flow rate exceeds the accumulated 
flow rate range.

Clear the accumulated flow rate.

System error

Displayed if an internal error has 
occurred.

Turn the power off and on again.

Alternately displays

[999] and [999999].

286

PFMC 

Series

Summary of Contents for PFMC Series

Page 1: ...Current PF2A 5 L min 1 L min Wide range of flow measurement with one product Compared with the current PF2A IP65 Compact Space saving Flow ratio 78 reduction Weight 74 reduction Setting resolution 10...

Page 2: ...dication shows the operating flow rate or residual amount of N2 etc in a gas cylinder Flow control of equipment main line and branch line Applications Remote control is possible with accumulated pulse...

Page 3: ...method Applicable fluid Thermal type MEMS Thermal type MEMS Air N2 Air N2 Argon CO2 Dry air N2 0 1 L min 0 5 L min 1 L min 2 L min 5 L min 5 L min 10 L min Thermal type MEMS Bypass flow type Thermal...

Page 4: ...o 50 C 12 to 24 VDC 10 12 to 24 VDC 10 1 to 10 L min 5 to 50 L min 10 to 100 L min 20 to 200 L min 50 to 500 L min 150 to 3000 L min 300 to 6000 L min 600 to 12000 L min PFMB Digital IP40 2 color LCD...

Page 5: ...5 M SI unit only Note 6 Note 5 Since the unit for Japan is fixed to SI due to new measurement law this option is for overseas Note 6 Fixed unit Instantaneous flow L min Accumulated flow L Calibration...

Page 6: ...25 to 525 L min Displays 0 when value is within 4 to 4 L min range 50 to 1050 L min Displays 0 when value is within 9 to 9 L min range 100 to 2100 L min Displays 0 when value is within 19 to 19 L min...

Page 7: ...a straight section of piping with a length of 8 cm or more If a straight section of piping is not installed the accuracy can vary by approximately 2 F S Straight section means a part of the piping wit...

Page 8: ...or solid state input for 30 msec or longer NPN 1 output Analog 1 to 5 V output type PFMC7 C NPN 1 output Analog 4 to 20 mA output type PFMC7 D Max applied voltage 28 V Max load current 80 mA Internal...

Page 9: ...ctifier Stainless steel 304 4 Sensor chip Silicon 5 Printed circuit board GE4F 6 Gasket HNBR 7 Body Aluminum alloy Anodized 8 Mesh Stainless steel 304 9 Spacer PPS 10 O ring HNBR 11 Holder Stainless s...

Page 10: ...ue Sheath Material Heat and oil resistant PVC Finished outside diameter 4 Symbol Model Port size A B D E F H K L N PFMC7501 7102 Rc1 2 NPT1 2 70 30 60 6 41 2 15 14 26 18 13 6 PFMC7202 Rc3 4 NPT3 4 G3...

Page 11: ...fixed state when starting the system or during maintenance This enables confirmation of the wiring and prevents system errors due to unexpected output For the analog output type when ON the output wi...

Page 12: ...ntents Action OUT1 over current error Load current of 80 mA or more is applied to the switch output OUT1 Eliminate the cause of the over current by turning off the power supply and then turn on it aga...

Reviews: