Chapter 3 Transport and Setting Up
HRSE Series
3.3
Installation
3-7
3.3.2
Electrical wiring
Power supply cable and Earth Leakage Breaker
Prepare the power supply shown in the following table. For the connection
between the product and power supply, use the power supply cable and
earth leakage breaker shown below.
Table 3-2 Power supply cable and Earth Leakage Breaker (
Recommended
)
Model
Power
supply
voltage
Power supply cable spec
Recommended earth
leakage breaker
Spec.
Size
Rated
current
Terminal
configuration
(to be onnected
to the user's
equipment)
Rated
voltage
Rated
current
Sensitivity
of leak
current
HRSE012-A-10
1-phase
100V AC
(50/60Hz)
3 cores x
14AWG
(3 cores x
2.0mm
2
)
(including
ground)
15A
Plug with ground
terminal
(JIS C8303 Plug
for the eceptacle
for 15A and 125 V
with dipole
grounding
electrodes)
Shared
with 100
V and
200 V
15A
15mA
or
30mA
HRSE012-A-10-T
HRSE018-A-10
HRSE018-A-10-T
HRSE012-A-20/23
1-phase
200V AC
(50/60Hz)
-23 Only
1-phase
230V AC
(50/60Hz)
3 cores x
14AWG
(3 cores x
2.0mm
2
)
(including
ground)
15A
Lead wire end not
treated
(Bare cut)
(L: White,
N: Black,
E: Green)
Shared
with 100
V, 200
V and
230V
15A
30mA
HRSE012-A-20/23-T
HRSE018-A-20/23
HRSE018-A-20/23-T
HRSE024-A-20/23
HRSE024-A-20/23-T
Only qualified persons are allowed to wire the product.
Be sure to shut off the user’s power supply. Wiring with the
product energized is strictly prohibitted.
The wiring must be secured to the product to prevent the external
force of cables being applied to the terminals. Incomplete wiring or
improper securing of wiring may cause electrical shock, excessive
heat and fire.
Ensure a stable power supply with no voltage surges.
Ensure that an Earth Leakage Breaker is used in the power supply
of the product. Se
Use a power supply suitable for the specifications of the product.
Be sure to connect the ground connection.
Ensure that a lock out facility is availble on the power supply.
Each product must have its own separate Earth Leakage Breaker.
Otherwise there can be a risk of electric shock or fire.
Do not modify the intenal electrical wiring of the product. Incorrect
wiring may cause electrical shock or fire. Also, modifing the
internal wiring will void the product’s warranty.
Do not connect the ground to water line, gas pipe or lightening
conductor.