PF2W5 20
03
T
PF2W504T
5 to 40
l
/min
0.5 to 4
l
/min
2 to 16
l
/min
PF2W520T
PF2W540T
Enclosure
Operating temperature range
Withstand voltage
Insulation resistance
Vibration resistance
Impact resistance
Noise resistance
Resistance
Wiring specifications
Remote type
Display unit
Flow rate range
04
20
40
0.5 to 4
l
/min
2 to 16
l
/min
5 to 40
l
/min
T
Temperature range
0 to 90
°
C
Nil
N
F
Rc
NPT
G
Thread type
Nil
1
2
Output specifications
Output for display unit
Output for display unit + Analog output (1 to 5 V)
Output for display unit + Analog output (4 to 20 mA)
Nil
N
3 m lead wire with connector
Without lead wire
Port size
Symbol
03
04
06
3/8
1/2
3/4
4
16
40
Applicable model
Flow rate (
l
/min)
PF2W504T/520T
PF2W520T/540T
PF2W540T
Port
size
Specifications
Water, Mixture of water (50%) and ethylene glycol (50%)
Karman vortex
0 to 1 MPa
1.5 MPa
0 to 90
°
C (With no cavitation)
±
5% F.S. or less
±
2% F.S. or less
±
2% F.S. or less (15 to 35
°
C based on 25
°
C),
±
3% F.S. or less (0 to 50
°
C based on 25
°
C)
12 to 24 VDC (Ripple
±
10% or less)
20 mA or less
IP65
Operating: 0 to 50
°
C, Stored: –25 to 85
°
C (No condensation or freezing)
1000 VAC for 1 min. between external terminal and case
50 M
Ω
(500 VDC) between external terminal and case
10 to 500 Hz at whichever is smaller: 1.5 mm amplitude or 98 m/s
2
acceleration in X, Y, Z directions for 2 hrs. each
490 m/s
2
in X, Y, Z directions 3 times each
1000 Vp-p, Pulse width 1
µ
s, Rise time 1 ns
1/2, 3/4
3/8, 1/2
3/8
660 g
Display units are the same as those of remote type digital flow switch for water (series PF2W3
).
Refer to page 16-13-19 for details.
Note 1) The system accuracy when combined with PF2W3
.
Note 2) Output system can be selected during initial setting.
Note 3) Without lead wire. (Add 20 g for the types of analog output whether voltage or current output selected.)
Note 4) The sensor unitis comformed to CE mark.
How to Order
Model
Measured fluid
Detection type
Flow rate measuring range
Operating pressure range
Withstand pressure
Operating fluid temperature
Linearity
Note 1)
Repeatability
Note 1)
Temperature characteristics
Power supply voltage
Current consumption (No load)
Weight
Note 3)
Port size (Rc, NPT, G)
Output
Note 2)
specifications
Output for display unit
Analog output
Pulse output, N channel, open drain, output for display unit PF2W3
.
(Specifications: Maximum load current of 10 mA; Maximum applied voltage of 30 V)
Voltage output 1 to 5 V within the flow rate range
Linearity:
±
5% F.S. or less; allowable load resistance: 100 k
Ω
or more.
Current output 4 to 20 mA within the flow rate range
Linearity:
±
5% F.S. or less; allowable load resistance: 300
Ω
or less with 12 VDC, 600
Ω
or less with 24 VDC
16-11-27
ZSE
ISE
PSE
Z
I
SE3
PS
Z
I
SE
1
2
ZSP
ISA2
IS
ZSM
PF2
IF
Data
Series
PF2W
Digital Flow Switch, High Temperature Fluid Type
For Water