7 PICOSCALE CONTROL GUI
7.7 Environmental Monitor
Figure 7.24: Environmental Monitor.
The
Environmental Monitor
displays the temperatur
e, humidity and pressure. When an environ-
mental module is connected to the
PICO
SCALE
and the GU
I
subscribes to the corresponding
event, temperature, humidity and pressure are recorded automatically. The in
fl
uence on the mea-
surement can be corrected using the
Deadpath Correction
for the individual channels.
I
n the center of the display you
fi
nd the environmental values plotted over time. Above the plot
the current values are shown. The environmental module records data with approximately one
Hertz sampling frequency and each plot can hold 24 hours of measurement data.
On the lower left hand side, the
Dead Path Correction
can be activated. During adjustment the
PICO
SCALE
roughly determines the absolute distance between target mirror and sensor head
and automatically determines the current dead path. The
Channel Dead Path [mm]
panels show
the dead paths (in millimeters) of the active channels.
Only when the dead path correction is enabled, variations of the refractive index are taken into
account and used for correction of position data. The dead paths can also be updated manually
when you enter a new value and enable the dead path correction for the speci
fi
ed channel.
When user de
fi
ned environmental parameters are inserted on the lower right side of the window
a new adjustment described in chapter 7.2 will be necessary if you want to switch back to standard
dead path correction. For further details on dead path correction, please refer to section 3.4.
76
PicoScale User Manual