Page 11
Figure 4. Wall Mounted Control Box
4. When the surgeon is ready to use the light,
install the sterilized center focus/positioning handle
using the following procedure. See figure 5. Be
sure handle is properly secured before using the
lighthead. Possible injury to patient or staff could
result if a handle is not installed properly.
a. Insert the handle into the lighthead at-
tachment ring.
WARNING
Be sure sterile handle is properly se-
cured before using the lighthead. An
improperly installed handle could fall
out, resulting in possible injury to pa-
tient or surgical staff.
b. Push the handle in, turn it right and left,
and pull the handle out to be certain that it is locked
(PUSH-TWIST-PULL). A distinct click can be heard
when the handle is properly engaged.
c. To remove the handle, push the release
button and pull the handle out.
CAUTION
Sterilizable handles are subject to nor-
mal wear and tear. Always examine the
handles for wear or damage to ensure
proper and safe operation with the
surgical light.
Figure 5. Center Focus/Positioning Handle
Installation
5. Adjust the focus by moving either the non-
sterile focus knob or the (sterilized) center focus
handle until all of the light beams converge on the
surgical site forming a single bright spot of light.
6. For low angle lighting approach, the lighthead
will move 90° below horizontal. Pull the lighthead
down by the positioning handles or the (sterile)
positioning/focus handle.
WARNING
Equipment not suitable for use in the
presence of a FLAMMABLE ANAES-
THETIC MIXTURE WITH OXYGEN OR
NITROUS OXIDE.
7. When the light is no longer required, return
the lighthead to its full up position. Decrease the
intensity at the wall control, and turn the main
power switch "OFF".
NOTE
SKYTRON products are guaranteed
for proper performance with the use of
genuine SKYTRON CENTER FOCUS
HANDLES. After market competitive
handles and other disposable handles
will have varying results that could ul-
timately affect the proper performance
and secure engagement of the center
focus handle. Such applications are
at the discretion of the user to ensure
patient safety.
SAFETY
FUSES
MAIN
POWER
SWITCH
INTENSITY
CONTROL
BUTTONS
COLOR
TEMPERATURE
CHANGE BUTTON
LED
CHANGE
INDICATOR
HANDLE
RELEASE
BUTTON
STERILIZABLE
HANDLE
Summary of Contents for AURORA LED 5 4000K
Page 1: ...REV 9 08 INSTALLATION INSTRUCTIONS LED SERIES SURGICAL LIGHTS...
Page 18: ...Page 16...
Page 20: ...Page 18...
Page 26: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...
Page 27: ...LED SERIES SURGICAL LIGHTS OPERATORS MANUAL TEC B 0002 REV0 1 09...
Page 47: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 48: ...PARTS CATALOG LED SERIES SURGICAL LIGHT TEC B 0009 REV3 3 10...
Page 54: ...Page 5 17 Stand Model Support Post Assembly page 40...
Page 65: ...Page 16 6 LED POD ASSEMBLY 57 010307 03mj 3 2 1 4 10 5 6 7 12 13 15 14 15 13 16 8 11 9...
Page 73: ...Page 24 10 LIGHTHEAD ASSEMBLY 1 2 6 7 8 9 11 13 14 12 10 5 3 4 57 031408 06mj...
Page 81: ...Page 32 13 TV LIGHTHEAD ASSEMBLY 1 3 4 5 6 7 8 9 10 11 12 15 17 18 13 16 14 2 5tv 081308 03...
Page 83: ...Page 34 14 WALL CONTROL ASSEMBLY 57 010307 04 1 5 6 9 10 3 2 7 8 4...
Page 85: ...Page 36 15 WALL CONTROL COMPONENTS 57 010307 05 1 2 3 4 5 6 12 13 14 7 8 9 10 11...
Page 91: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 92: ...MAINTENANCE MANUAL LED SERIES SURGICAL LIGHT 7 07...
Page 93: ......
Page 95: ...Page 2...
Page 110: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 111: ...REV 8 08 INSTALLATION INSTRUCTIONS LFS SERIES SURGICAL LIGHTS...
Page 133: ...Page 21...
Page 135: ...Page 23...
Page 138: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...
Page 139: ...LFS SERIES OWNERS MANUAL REV 6 08...
Page 140: ......
Page 142: ...Page 2...
Page 160: ...Page 20...
Page 170: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...
Page 171: ...INSTALLATION INSTRUCTIONS AUR SERIES SURGICAL LIGHTS TEC B 0006 REV1 8 09...
Page 188: ...Page 16 TYPICAL AURORA AUR LIGHT FIXTURE CONDUIT AND WIRING REQUIREMENTS...
Page 189: ...Page 17 TYPICAL WIRING DIAGRAM...
Page 190: ...Page 18 TYPICAL AURORA AURTV LIGHT FIXTURE CONDUIT AND WIRING REQUIREMENTS...
Page 191: ...Page 19 TYPICAL WIRING DIAGRAM...
Page 196: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...
Page 197: ...AUR SERIES SURGICAL LIGHTS OPERATORS MANUAL TEC B 0003 REV9 3 10...
Page 198: ......
Page 219: ......
Page 220: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 221: ...PARTS CATALOG AUR SERIES SURGICAL LIGHT TEC B 0004 REV6 6 10...
Page 222: ......
Page 274: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 275: ...AUR SERIES SURGICAL LIGHTS MAINTENANCE MANUAL TEC B 0007 REV1 1 10...
Page 276: ......
Page 278: ...Page 2...
Page 292: ...Page 16 TYPICAL WIRING DIAGRAM...
Page 293: ......
Page 294: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 295: ...TEC B 0008 REV3 3 10 OWNERS MANUAL HANDLE CAMERA SYSTEM...
Page 296: ......
Page 308: ...Page 12 PRECISION HD CAMERA SYSTEM...
Page 310: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...