Installation and operating instructions AccuFlo

®

Zero 

 

9  Technical data 

9.1 

General information 

Automatic zero-point calibration 

 

Cycle length between calibrations 

Configurable at order (factory setting: 24 h) 

Temperature trigger 

Configurable at order (factory setting: 10 K 
temperature difference between last calibration) 

Length 

30 s 

Maintaining of analog output signal 

Last available measurement output value is held 
constant by transmitter during zero-point calibration  

Commissioning mode 

Correction of permanent zero-point errors due to 
installation errors   

Supported flow meters  

All differential pressure based flow meters (ISO 5167 
and AGA-3-primary elements, pitot tubes, v-cones, 
etc.) 

Process and signal connection

 

 

Material of wetted parts 

Stainless steel (1.4571) 

Process connection manifold 

IEC 61518 / DIN EN 61518 Groove Ø18.5 mm 

Process connection transmitter  

IEC 61518 / DIN EN 61518 Type A with spigot 

Pressure rating 

PN 100 

Ambient conditions

 

 

Ambient temperature of wetted parts 

–20 … 50 °C 

Climate conditions 

Humidity 0

–100 %, external moisture condensation of 

installed device permitted.  

Protection class 

IP 65 ( IEC 60529) 

Electromagnetic protection 

According to EN55011:2009+A1:2010, Group 1, 
Class B und EN61326-1:2013 

Design

 

 

Weight 

2.8 kg 

Dimensions 

120x40x140 mm³ 

Life expectancy 

At least 10.000 calibrations 

Retrofittability 

For all differential pressure based flow meters 

Certificates and standards

 

 

Pressure Equipment Directive  

DGRL 2014/68/EU 

For gases fluid group 1, liquids fluid group 1; 
fulfills requirements according to Art. 4 (3) (good 
engineering practice) 

Summary of Contents for AccuFlo Zero

Page 1: ...AccuFlo Zero Automatic Zero Point Calibration For Differential Pressure Transmitters Installation and Operating Instructions...

Page 2: ...ctions 8 5 3 Status relay 9 5 4 Manual zero point calibration 9 5 5 Terminal assignment 9 6 Commissioning 13 6 1 General procedure 13 6 2 Successful set up 13 6 3 Optional correction of positioning wi...

Page 3: ...ce may lead to serious injury and or considerable damage to property 1 2 Qualified personnel The installation and operating must be realized by qualified personnel Qualified personnel includes persons...

Page 4: ...es of the product are not opened under pressure Fire near the product Appropriate measures are to be taken to secure that the product is taken out of operation in case of damage Improper mounting of t...

Page 5: ...l if the transmitter supports the necessary functions via HART The automatic zero point calibration is executed directly in the transmitter analog to a manual calibration with a possible present local...

Page 6: ...f supply for the following items AccuFlo Zero AccuFlo Zero 1 AccuFlo Zero 2 Threaded rods and screw nuts 7 16 UNF or M10 according to the transmitter for assembling of the system 3 PTFE seals for the...

Page 7: ...n case of measurement of liquids or steam the two locking screws without valve see above in Figure 1 Assembled AccuFlo Zero are to be replaced by the delivered locking screws with valve bleeder valve...

Page 8: ...cribed Power supply o The AccuFlo Zero needs a voltage supply of 24 V DC with a power rating of at least 2 1 ampere o We recommend a switch mode power supply o We recommend positioning the power suppl...

Page 9: ...ent loop is fixed and shown statically during this period Both relays may be connected in series in the AccuFlo Zero and provide a group signal A 1 signal is only sent if the AccuFlo Zero is switched...

Page 10: ...inal Terminal in transmitter Use A4 Differential pressure transmitter 4 20 mA D2 Connection between the AccuFlo Zero and the control station Table 3 Analog signal to the control station Terminal Descr...

Page 11: ...tputs option 3 Terminal Description Use E2 external Externally supplied signal E3 Calibration Signal Calibration of the manually supplied signal after successful zero point calibration E4 Calibration...

Page 12: ...Zero If necessary external triggering of the zero point calibration Table 8 Signal start zero point calibration from the control station to the AccuFlo Zero Terminal Description Use D4 GND GND of the...

Page 13: ...in the pipe system is stopped and there is no flow 2 Ensure that the system differential pressure transmitter AccuFlo Zero manifold impulse pressure lines and the condensate vessels are filled with wa...

Page 14: ...sult without positioning correction 6 5 Signaling to the control station Status information may be transmitted from the AccuFlo Zero to the control station Two relays are available For more informatio...

Page 15: ...entional power supply with classical transformer A substitution is recommended o Shielded wires may decrease the influence of external interference Connect the shield only one sided with the ground po...

Page 16: ...elements pitot tubes v cones etc Process and signal connection Material of wetted parts Stainless steel 1 4571 Process connection manifold IEC 61518 DIN EN 61518 Groove 18 5 mm Process connection tran...

Page 17: ...0 mA Supply of the transmitter Provided Digital communication HART Maximum wire length between AccuFlo Zero and transmitter 5 m Outputs Output signal Two wire interface 4 20 mA Maximum wire length bet...

Page 18: ...AccuFlo Zero Installation and Operating Instructions 18 Installation and operating instructions AccuFlo Zero 10 Declaration of Confirmity...

Page 19: ...AccuFlo Zero Installation and Operating Instructions 19 Installation and operating instructions AccuFlo Zero...

Page 20: ...on GmbH Postfach 41 01 31 D 41241 M nchengladbach Hanns Martin Schleyer Str 22 D 41199 M nchengladbach Tel 49 0 2166 62317 0 Web www ski gmbh com E Mail info ski gmbh com Registered trademarks and log...

Reviews: