AccuFlo
®
Zero
– Installation and Operating Instructions
11
Installation and operating instructions AccuFlo
®
Zero
Option 2:
+24-V-signal from AccuFlo
®
Zero to the control
station mode as group message
Remove/
open jumper „J2“. Connect terminal “E4” with “F4”!
Table 5: Relay output - option 2:
Terminal Description
Use
E3
+
Signal „Ready and Calibration Successful”
+24 V if AccuFlo
®
Zero is ready and zero-
point calibration is successful
Option 3: Externally supplied signal from AccuFlo
®
Zero to
control station for
status „Error“ and „Ready“
Remove/
open jumper „J2“.
Table 6: Relay outputs
– option 3:
Terminal Description
Use
E2
ex *
Externally supplied signal
E3
+ (Calibration)
Signal „Calibration“
+ of the manually supplied signal after
successful zero-point calibration
E4
+ (Calibration)
Signal „Ready“
+ of the manually supplied signal if the
AccuFlo
®
Zero is ready
Option 4: Externally supplied signal from AccuFlo
®
Zero to
control station as “group message“
Remove/open
jumper „J2“.
Table 7: Relay outputs
– option 4:
Terminal Description
Use
E4
ex *
Externally supplied signal
E3
+ (Ready and
calibration
successful)
Signal „ Ready and calibration successful “
+ of the manually supplied signal, if the
AccuFlo
®
Zero is ready and the zero-point
calibration is successful
*Since mechanical relays may be used, also for example
alternating voltage may be chosen.