Run Time Features:
User-selectable features, user-adjustable presets (
see
Advanced
Setup
)
Preset timed runs:
1, 2, 5, 10, or 15 min
Manually set continuous
run with manual shut-off:
1 to 999 min (repeats 1 to 999-min
runs
inde
fi
nitely until user stops
pump or power supply is depleted)
Intermittent
sampling:
See Advanced Setup
Flow Fault Features:
If the pump is unable to compensate due to excessive back
pressure, the pump will go into
fl
ow fault mode
(see page 18 for
details)
. Flow fault features are enabled as the default setting.
Flow fault may be disabled at the user’s option.
See Advanced Setup
.
Flow
fault:
Flow fault displays immediately
Pump shuts off after 5 seconds
Fault
restart:
Pump attempts restart every 10 seconds
up to
fi
ve times
Media Compatibility:
Viable cascade impactors, spore trap cassettes (e.g., VersaTrap),
asbestos cassettes, microvacuum cassettes, and other impaction
samplers that require
fl
ows from 10 to 30 L/min
Tubing:
Requires 3/8-in ID tubing
Housing:
ABS
plastic
Dimensions:
9.3 x 8.4 x 3.5 in (23.6 x 21.3 x 8.9 cm)
Weight:
4.8 lbs (2.2 kg)