![SKC QuickTake 30 Operating Instructions Manual Download Page 5](http://html1.mh-extra.com/html/skc/quicktake-30/quicktake-30_operating-instructions-manual_1286857005.webp)
4
Performance Pro
fi
le
Flow Range:
10 to 30 L/min
Flow Compensation
Accuracy:
± 5% of set
fl
ow
Compensating Flow
10 L/min at 90 inches water back pressure
Back Pressure Range:
20 L/min at 50 inches water back pressure
30 L/min at 15 inches water back pressure
Power:
•
Rechargeable lithium-ion (Li-Ion) battery,
7.4 V, 9.6-Ah
capacity, 71.04 Wh
•
Battery with AC charger/adapter,
100-240 V
Battery Recharge Time:
Approximately 5 hrs
(varies with ba
tt
ery capacity
and level of discharge)
Typical Cumulative
Spore Trap*
(e.g., VersaTrap)
:
5 hrs at 15 L/min (
battery only
)
Run Time:
†
BioStage viable cascade impactor:
4 hrs at 28.3 L/min (
battery only
)
25-mm, 1.2-
μ
m MCE
fi
lter:
9+ hrs at 10 L/min
To achieve, use QuickTake 30 with a fully
charged battery
and
AC charger/adapter.
37-mm, 0.8-
μ
m MCE
fi
lter:
> 14 hrs at 10 L/min (
battery only
)
* Sampling times when using spore traps are usually
≤
10 min. SKC
recommends reducing length of sample time when using a 30 L/min
fl
ow rate to prevent overloading the media.
† Results obtained using a new pump and new fully charged ba
tt
eries. Pump and
ba
tt
ery performance may vary.
The AC charger/adapter can be used to extend ba
tt
ery run
time, but it is not a ba
tt
ery eliminator. Therefore, it will
not provide inde
fi
nite run times.
Temperature:
Operating:
32 to 113 F (0 to 45 C)
Storage:
-4 to 95 F (-20 to 35 C)
Charging:
32 to 113 F (0 to 45 C)
Operating Humidity:
0 to 95% non-condensing
Protect sample pump from weather when in use outdoors.
Compare the information in this table to pump compensation range to determine appropriate applications.
Typical Back Pressure of Sampling Media (
inches water
)
Flow Rate (L/min)
10
12
15
20
Filter/Pore Size (
μ
m)
25-mm MCE, 0.8
65
80
103
148
25-mm MCE, 0.45
138
172
> 200
—
37-mm MCE, 0.8
22
28
36
51
37-mm PVC, 5.0
9
11
15
21