Instructions for use

1

The device may only be used for purposes

approved by the manufacturer

Before the actual measurement, it is necessary

to perform the calibration, see: use

Perform calibration regularly

Make sure the prism and lid are clean

Do not clean with corrosive or abrasive

substances

Do not rinse the device with water, only carefully

clean the prism (with the enclosed cloth) to

prevent water from entering under the prism

Any damage to the prism will damage the device

For measurement accuracy, the prism and test

When testing the battery electrolyte, avoid

contact with the skin or eyes !!!

Notice

specimen must have the same temperature

Use

1. Open the cover

2. Pipette 2-3 drops of distilled water on the prism

3. Close the cover so that the liquid can be evenly

distributed over the entire surface of the prism to

prevent the formation of dry spots / bubbles

4. Wait 30 seconds for the temperatures to equalize

5. Aim the cap at the light source and look into the

eyepiece

6. You will see a circular field with several scales

7. Rotate the eyepiece to focus the scale

8. The upper part of the circular field should be blue

9. The bottom should be white

10. Turn the calibration screw until the boundary

between the top and bottom reaches 0 (middle

scale).

Custom measurement

1. Apply a few drops of the test sample to the prism

2. Close the cover

3. Aim it at the light source and look into the

eyepiece

4. The concentration value is determined on the

scale by the place where the blue and white

boundaries are shifted

Do not clean with corrosive or abrasive

substances

Maintenance, cleaning

Do not rinse the device with water, just carefully

clean the prism (supplied cloth) to prevent water

from entering the prism

Alcohol, in particular wine, vodka, whiskey and

other destilates

Measurement options

EN

Summary of Contents for SX3003

Page 1: ...Refractometer EN Refraktometr CZ Refraktometer SK SX3003 Instruction for use N vod k pou it N vod na pou itie...

Page 2: ...surface of the prism to prevent the formation of dry spots bubbles 4 Wait 30 seconds for the temperatures to equalize 5 Aim the cap at the light source and look into the eyepiece 6 You will see a cir...

Page 3: ...ion legislation of the European Union EMC 2014 30 EU This declaration is issued under the sole responsibility of the manufacturer Harmonized standards including their amending annexes if any which hav...

Page 4: ...ut vody pod prizma Pou it 1 Otev ete v ko 2 Naneste pipetou 2 3 kapky destilovan vody na prizmu 3 Uzav ete v ko tak aby se kapalina mohla rovnom rn rozprost t po cel plo e prizmy aby se nevytvo ila su...

Page 5: ...r vn mi p edpisy Evropsk unie EU 2011 65 EU 2014 30 EU 2014 35 Toto prohl en se vyd v na v hradn odpov dnost v robce Harmonizovan normy v etn jejich pozm uj c ch p loh pokud existuj kter byly pou ity...

Page 6: ...trie po celej ploche prizmy aby sa nevytvorili such miesta bubliny 4 Vy kajte 30 sek nd na vyrovnanie tepl t12 5 Nasmerujte vie ko k zdroju svetla a pozrite sa do okul ra 6 Uvid te kruhov pole s nieko...

Page 7: ...vnymi predpismi Eur pskej nie EU 2011 65 EU 2014 30 EU 2014 35 Toto prehl senie sa vyd va na v hradn zodpovednos v robcu Harmonizovan normy vr tane ich pozme uj cich pr loh pokia existuj ktor boli po...

Page 8: ...jeme si pr vo u init zm ny v tomto n vodu k pou it bez p edchoz ho upozorn n Nejaktu ln j verze n vodu k pou it je k dispozici na eshopu v kart tohoto produktu SK Vyhradzujeme si pr vo urobi zmeny v t...

Reviews: