60 90 570 D 3485
28
D 3485.201.04.09.02
09.2007
5 Operating Instructions
5
Operating Instructions
CAUTION
Always protect the optical fiber as well as its connector with the special protec-
tion caps at the end of treatment (see Fig. 17, page 24).
5.1
Warning and error messages
After selecting the indication, press the “ENTER” key. The laser unit is now
ready for operation. The message “Wear protective goggles” appears on the
display and four green LEDs start flashing.
If this message is displayed, check whether the optical fiber is properly con-
nected. Once the optical fiber is properly connected, the message will dis-
appear automatically.
If this message is displayed, you must select an optical fiber diameter with
the “RIGHT ARROW” key and then confirm your selection with the “ENTER”
key.
If this message is displayed, you must insert the specified optical fiber and
then confirm your action with the “ENTER” key.
This message signals that the INTERLOCK safety device has responded.
Either the door of the treatment room is open, or the INTERLOCK connector
is not properly seated in its socket.
If this message is displayed, allow the laser unit to cool down for a few min-
utes.
The temperature of the laser diode is
< 5 °C or > 90 °C; or
the temperature
sensor is defective.
Adapt the SIROLaser to room temperature over a period of 2 hours.
If this message nevertheless still appears, contact your dealer or an authorized
Customer Service Department.