SAFETY STANDARDS AND CLASSIFICATION OF THE SYSTEM
SICAT JMT
⁺
Jaw Motion Analysis System
Page 31 of 37
GUIDELINES AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC INTERFERENCE
IMMUNITY
The measurement systems in the JMT range of products are intended for operation in the
electromagnetic environment stated below. The customer or user of the JMT measurement system
should ensure that it is used in such an environment.
Immunity tests
IEC 60601- test
level
Compliance level Electromagnetic environment -
guidelines
Electrostatic
discharge (ESD)
according to IEC
61000-4-2
± 6 kV contact
discharge
± 8 kV air
discharge
± 6 kV contact
discharge
± 8 kV air
discharge
Floors should be made from wood or
concrete or fitted with ceramic tiles. If
the floor is equipped with synthetic
material, the relative air humidity
should be at least 30%.
electrical fast
transient/bursts
according to IEC
61000-4-4
± 2 kV for power
cables
± 1 kV for input
and output cables
± 2 kV for power
cables
± 1 kV for input
and output cables
The supply voltage quality should be
that of a typical commercial or hospital
environment.
Surges according
to IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
The supply voltage quality should be
that of a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on the supply
voltage according
to IEC 61000-4-11
< 5% U
T
(> 95% drop to
the U
T
) for ½ a
period
40% U
T
(60% drop
to the U
T
) for 5
periods
70% U
T
(30% drop
to the U
T
) for 25
periods
< 5 % U
T
(> 95 %
drop to the U
T
) for
5 s
< 5% U
T
(> 95%
drop to the U
T
) for
½ a period
40% U
T
(60% drop
to the U
T
) for 5
periods
70% U
T
(30% drop
to the U
T
) for 25
periods
< 5 % U
T
(> 95 %
drop to the U
T
) for
5 s
The supply voltage quality should be
that of a typical commercial or hospital
environment. If the user of the
measurement systems in the JMT
range of products requires continued
functioning during power supply
interruptions, it is recommended that
the JMT jaw motion analysis system be
powered by an interruption-free power
supply or a battery.
Magnetic field
with the power
supply frequency
(50/60 Hz)
according to IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should correspond with the typical
values that can be found in the
commercial or hospital environment.
NOTE U
T
is the A.C. mains voltage prior to application of the test level.