FUNCTIONAL CHECKS, PREPARATION, DISPOSAL
Page 26 of 37
SICAT JMT
⁺
Jaw Motion Analysis System
MANUAL DISINFECTION
The measurement system can be wipe-disinfected with suitable solutions. Disinfect all components with
a cloth that has been dampened with a disinfectant solution.
CAUTION
Do not use any spray disinfectants on the head bow or the LJ-sensor!
Spray disinfectants destroy the high accuracy measurement sensors.
RECOMMENDED DISINFECTANT SOLUTION
Composition approx. 25% ethanol, 35% propanol
e.g. "Mikrozid "Liquid / Schülke & Mayr or comparable solutions
NOTICE
When using a disinfectant solution, please comply with the recommendations
stated by the manufacturer, especially the specified application time.
WARNING
Due to the risk of possible mix-ups, chemicals that are required for the
disinfection or cleaning must only be stored, prepared and kept ready in their
intended containers.
STERILIZATION
All accessory parts that come into contact with the patient's mucosa have to be sterilized before use.
NOTICE
The sterilization is to be completed immediately subsequent to the cleaning.
Sterilize the bite-forks and lower jaw attachment with a fractionated pre-vacuum for four minutes at
134 °C ± 1 °C and 3.04 bar (can be sterilized up to a max. of 138 °C).
It is necessary to sterilize the following accessory parts:
■ Para-occlusal attachment 90
■ Occlusal adapter
■ Occlusal attachment
■ Bite fork type SI
DISPOSAL
PACKAGING
In Germany, all of the transport packaging supplied by SICAT can be disposed of into the materials cycle
at local waste disposal centers.