7 Service
Operating Instructions CEREC MC XL Premium Package
7.5 Care, cleaning agents, and disinfectants
båÖäáëÜ
10. Close the processing chamber door.
11. In the software, select which instrument(s) you have used and click
on
"Start"
.
You can also select the instruments on the touch display of the unit
and confirm with
"Start"
.
Changing a defective instrument
If an instrument breaks during the machining phase, the corresponding
motor travels to the change position. A dialog box which marks the side
with the broken instrument with a red cross then opens.
✔
The instrument is broken.
1. Change the defective instrument as described above.
2. Select the instrument which you have inserted.
3. Press the
"Start"
button.
7.5
Care, cleaning agents, and disinfectants
A continuously updated list of approved media can be downloaded from
the internet via the online portal for technical documents. You can reach
this portal at the address:
www.dentsplysirona.com/manuals
. Click on the menu item
"General documents"
and then open the
"Care,
cleaning and disinfection agents"
document
If you do not have internet access, please contact your dental depot to
order the list (REF 59 70 905).
NOTICE
Incorrect milling or grinding results
Interchanging instruments leads to faulty milling or grinding results.
NOTICE
Cleaning cooling water nozzles
The cooling water nozzles in the processing chamber must be free from
limescale, grinding dust/milling dust deposits. The corresponding
cooling water jet must always strike the instrument accurately!
✔
The cooling water nozzles are dirty.
➢ Clean the nozzles with a cleaning wire and a syringe.
NOTICE
Use only suitable instruments!
Do not use any instruments from CEREC 2 or CEREC 3 units.
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents approved by Dentsply
Sirona!