Operating Instructions CEREC MC XL Premium Package
Inserting instruments
1. Open the processing chamber door following calibration.
2. Loosen the calibration pins with the torque wrench and remove them.
3. Loosen the ball pressure screw.
4. Remove the calibration phantom by grasping it by its narrow surfaces
(B).
5. Insert the instruments in the motor mount by hand. Tighten the
corresponding chuck with the torque wrench until a clicking sound
can be heard.
6. Close the processing chamber door.
The dialog box for selecting the instruments then appears.
7. Select the inserted instruments and confirm by clicking the
"Start"
button in the window on the PC or touch display.
The motor mounts move to the initial position.
The
"Calibration succeeded"
dialog window appears.
Exiting the calibration
1. Click the
"OK"
button on the PC or on the touch display.
2. Click the
"Exit Configuration"
step.
6.3
Machining process
6.3.1
Process types
Various process types are available for milling and grinding purposes.
These vary in terms of the type of materials to be processed and the
instruments to be used.
NOTICE
Store the calibration tools in a safe place
Store the calibration pins and the calibration body in a safe place (e.g.
in a storage box in the unit drawer).