3 Vacuum system
Sirona Dental Systems GmbH
3.5 Cleaning the suction hoses
59 06 511 D 3386
12
D 3386.103.01.06.02
3.5
Cleaning the suction hoses
The hoses of the saliva ejector and suction cannula as
well as the connection hose to the water unit can be
pulled off from the assistant element for rinsing under
running water or thermodisinfection.
Switch the unit OFF first!
Unscrew the suction hoses from the top parts (5) of the
filter housings. Rinse the hoses under running water.
The top parts of the housings (5) with filter inserts must
be thoroughly cleaned every day.
The top parts of the filter housings can be
thermodisinfected.
Remove the amalgam residue collectors (6) from the
lower parts (7) if the suction strength decreases,
otherwise once a day.
Store amalgam residues under water in a separate
container. When the amalgam rotor is replaced (see
page 14), put the residues into the rotor for disposal.
ATTENTION
Amalgam residues must not be disposed of in the
cuspidor or the sink!
When the surfaces of the hoses have become sticky
from disinfectants, clean them with regular dish washing
fluid and then rub talcum over the hoses, if necessary.
Before refitting the hoses, grease the O-rings at the
connection pieces.
ATTENTION
Use only lubricants which have been approved by Sirona
(see page , General Notes)!
The hoses must engage firmly.
If the unit is equipped with a 3rd suction hose, clean it the
same way.
5
7
6
Grease
Summary of Contents for C1+
Page 2: ......
Page 4: ......
Page 5: ... NH mÑäÉÖÉ ìåÇ oÉáåáÖìåÖ ÇìêÅÜ Ç ë mê ñáëíÉ ã aÉìíëÅÜ ...
Page 8: ...Inhaltsverzeichnis Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 01 ...
Page 33: ......
Page 35: ... NH êÉ åÇ äÉ åáåÖ Äó íÜÉ mê ÅíáÅÉ qÉ ã båÖäáëÜ ...
Page 38: ...Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 02 ...
Page 63: ......
Page 65: ... NH båíêÉíáÉå Éí åÉííçó ÖÉ é ê äÛ èìáéÉ Çì Å ÄáåÉí cê å áë ...
Page 68: ...Sommaire Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 03 ...
Page 93: ......
Page 95: ... NH ìáÇ Çç ó äáãéáÉò éçê é êíÉ ÇÉä éÉêëçå ä ÇÉ ä Åçåëìäí bëé çä ...
Page 98: ...IÍndice Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 04 ...
Page 123: ......
Page 125: ......