Sirona Dental Systems GmbH
3 Vacuum system
3.4 Suction handpieces
59 06 511 D 3386
D 3386.103.01.06.02
11
båÖäáëÜ
3.4
Suction handpieces
Sterilization
The suction cannula (1),
the intermediate piece (2),
the suction handpiece (3),
the saliva ejector handpiece (4)
and the surgical suction cannula (5)
can be sterilized.
Sterilize only in an autoclave at 135
°
C, 2.1 bar.
Thermodisinfection
All parts can be thermodisinfected at 93
°
C (197°F).
Disinfection
All parts can be sprayed or wiped with a disinfecting
solution.
ATTENTION
Use only cleaning agents which have been approved by
Sirona !
Grease the O-rings at the joints afterwards!
Weekly
Separate the suction handpiece at the joint, clean the
handpiece, and grease the joints.
ATTENTION
Use only lubricants which have been approved by
Sirona!
1
2
3
4
5
3
Grea
se
Grease
Summary of Contents for C1+
Page 2: ......
Page 4: ......
Page 5: ... NH mÑäÉÖÉ ìåÇ oÉáåáÖìåÖ ÇìêÅÜ Ç ë mê ñáëíÉ ã aÉìíëÅÜ ...
Page 8: ...Inhaltsverzeichnis Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 01 ...
Page 33: ......
Page 35: ... NH êÉ åÇ äÉ åáåÖ Äó íÜÉ mê ÅíáÅÉ qÉ ã båÖäáëÜ ...
Page 38: ...Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 02 ...
Page 63: ......
Page 65: ... NH båíêÉíáÉå Éí åÉííçó ÖÉ é ê äÛ èìáéÉ Çì Å ÄáåÉí cê å áë ...
Page 68: ...Sommaire Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 03 ...
Page 93: ......
Page 95: ... NH ìáÇ Çç ó äáãéáÉò éçê é êíÉ ÇÉä éÉêëçå ä ÇÉ ä Åçåëìäí bëé çä ...
Page 98: ...IÍndice Sirona Dental Systems GmbH 59 06 511 D 3386 4 D 3386 103 01 06 04 ...
Page 123: ......
Page 125: ......