Safety Issues
CAUTION: Be Sure to Disconnect the CDR PlusWire Sensor from the CDR
Remote, or CDR Remote HS, or SDX Sensor Interface before Performing
Replacement Procedures
To avoid potential damage to CDR PlusWire Sensor components
during cable replacement, always disconnect the Sensor and its
cable from the CDR Remote, or CDR Remote HS, or SDX Sensor
Interface. When performing cable replacement, always work outside
the patient area, using the tools and materials supplied by Sirona
Dental.
CAUTION: Be Sure to Disconnect the CDR PlusWire Sensor from the CDR
Remote, or CDR Remote HS, or SDX Sensor Interface before Performing
Cleaning and Disinfecting Procedures
To avoid potential damage to CDR PlusWire Sensor components
during cleaning and disnfecting, always disconnect the Sensor and
its cable from the CDR Remote, or CDR Remote HS, or SDX
Sensor Interface.
Check CDR PlusWire Sensors before Using Them
Before each usage, check the outer surface of the CDR PlusWire
Sensor for any signs of physical damage or defect. The surface of
the CDR PlusWire Sensor should have a smooth finish, with no
evidence of chipping or damage. If detected, discontinue use of that
sensor and contact your local distributor of Sirona Dental products
for further instructions.
The CDR PlusWire cable should be checked for any damage in the
shielding, such as exposed or frayed shields or broken connections
to solder points. If detected, discontinue use of that cable and
contact your local distributor of Sirona Dental products for further
instructions.
Always Use Sheaths with CDR PlusWire Sensors
Use Sirona Dental positioning products and sheaths every time the
CDR Plus Wire Sensor is used.
Never use the Sensor without a
protective sheath
. Never use a damaged sheath.
Always dispose of the sheath after every patient. Protective sheaths
are single-use devices and must not be reused under any
circumstance. Reuse of single-use devices/instruments may cause
them to become contaminated, compromise their intended function,
and result in patient and user infection, injury and/or illness.