B1051067 Rev. A
CDR PlusWire User Guide
10
3.3. Description of the CDR PlusWire Utility
The Sensor edge-card does more than just provide the means for the
CDR Sensor to connect to USB Remotes or to the Sensor interface
on the SDX system. It also includes specific information about the
Sensor itself and uniquely identifies it for the purpose of Sensor
calibration.
By replacing the Sensor cable, Sensor-specific information could be
lost without the ability — supplied by the CDR PlusWire Utility —
to read that information from one edge-card connector, and write it
to the new edge-card connector. To accomplish this, the utility
consists of a few dialog boxes that will prompt you through the
entire process.
Refer to Section 3.5 on page 11 for step-by-step directions on cable
replacement for the CDR Plus Wire sensor.
3.4. After Replacing the CDR PlusWire Cable
Any additional items that were not used during the replacement
procedure (extra labels, for example, or elastomeric strips) can be
retained.
International customers residing in countries where the Waste
Electrical and Electronic Equipment (WEEE) Directive has been
implemented should contact their local dealers for more information
regarding the proper disposal of the failed cable assembly.