28
Figure 15 CCIR17 Waveform
2T20TPB (960H only)
The 2T20TPB signal is a pulse, bar and two pulses, 2T and 20T wide. It
allows the measurement of K-factor and luma/chroma gain and delay. A
separate signal to the CCIR17 and NTC-7 waveforms is created because
960H has a wider bandwidth (e.g. 18MHz luma sampling, compared to
13.5MHz luma sampling for NTSC/PAL).
Black
Black is a full frame 0 IRE (NTSC-M) or 0mV (PAL) video signal that may
be used for noise measurements.
White
White is a full frame 100 IRE (NTSC-M) or 700mV (PAL) video signal that
may be used for noise measurements or for finding AC coupling or
clamping issues.
Pedestal (525 line), 50% F/F (625 line)
Pedestal is a full frame 50 IRE (NTSC) or 350mV (50%, PAL) video signal
that may be used for noise measurements. The noise measurement of the
625 Line CCIR17
0
100
200
300
400
500
600
700
800
900
1000
0
100
200
300
400
500
600
700
800
Y Pixel No.
1
0
b
it
v
a
lu
e
0
100
200
300
400
500
600
700
800
900
0
50
100
150
200
250
300
350
400
Cb/Cr Pixel No.
Y Value (Dec)
Cb Value (Dec)
Cr Value (Dec)
Summary of Contents for SM01
Page 12: ...12 Menu control The complete menu structure is shown in Figure 4 Figure 4 SM01 Menu structure ...
Page 44: ...44 Figure 30 PAL vertical timing composite output ...
Page 45: ...45 Figure 31 NTSC Horizontal timing composite output ...
Page 47: ...47 Figure 33 NTSC 75 colour bar composite output ...
Page 48: ...48 Figure 34 NTSC 75 colour bars vectors ...
Page 49: ...49 Figure 35 PAL 75 colour bars composite output ...
Page 53: ...53 Figure 39 NTSC Differential gain phase measurement ...
Page 54: ...54 Figure 40 NTSC noise measurement Pedestal waveform ...
Page 55: ...55 Figure 41 Luma frequency response CCIR18 test signal ...
Page 58: ...58 Figure 45 SDI output 625 line eye pattern ...
Page 59: ...59 Figure 46 SDI output 625 line jitter measurement ...
Page 61: ...61 Figure 48 AC DC converter specification Page 2 ...
Page 62: ...62 Figure 49 AC DC converter specification Page 3 ...