119
GB
Error
Problem
Solution
“Hi” on the
LC display
4
(measured
value above
42.9°C or
109.2°F)
Check the meas-
urement mode.
Please observe
the directions
for use with re-
gard to taking
correct measure-
ments.
“ErrE” on the
LC display
4
Ambient tempera-
ture outside the
range of 15.0 °C
- 40.0 °C /
59.0 °F -
104.0 °F (fore-
head mode) and
10.0 °C -
40.0 °C /50.0 °F
- 104.0 °F (ear
and scan mode).
Please ensure an
ambient tempera-
ture in the range
of 15.0 °C -
40.0 °C /
59.0 °F -
104.0 °F (fore-
head mode) and
10.0 °C -
40.0 °C/
50.0 °F -
104.0 °F (ear
and scan mode)
“ErrU” on the
LC display
4
No measure-
ment result after
measurement.
Please observe
the directions
for use with re-
gard to taking
correct measure-
ments.
“ErrH” on the
LC display
4
Error detected
during self-test.
Contact the
service centre.
“ErrP” in LC
display
4
Hardware
problem
Contact the
service centre.
IB_106635_SSOT6B2_DE_FR_IT_GB_LB1.indd 119