5 Commissioning
The following diagrams apply for work execution (unscaled, see user man-
ual).
Following mounting and complete wiring, the system can be commis-
sioned. The sensor is factory-precalibrated.
5.1 Analog voltage output
The internal resistance of the downstream electronics unit must be >5 kΩ.
Connect the appropriate open ends of the connection cable with your mea-
suring system.
1 axis: pins 5 and 4, see chapter
2 axes: for X axis pins 5 and 4, for Y axis pins 7 and 4, see chapter
1
Output Z-axis
2
Follower electronic
Fig. 3: Analog voltage output
00.00 V
GND
+
1
2
1 axis
Conversion of voltage (Uout) to the angle:
Position angle (in °) = (10 V) / (0.02777 Volt pro °)
Example: Uout = 1.6662 V
Position angle = (1.6662 V) / (0.02777 V pro °) = 60°
5
0
45
90
135
180
225
270
315
359.9
10
0
Volt
age (V)
Angle (°)
NOTICE
NOTICE