Operating the Calibrator / Micro Calibration Bath
7.3.4.3 Setting the duration time (S version)
The duration time
"dur.t"
is active if the set temperature SP1 has been achieved. Subse-
quently, the calibrator / micro calibration bath automatically switches to set temperature SP2.
The setting range extends from 99:59 [hh:min] to 00:00 [hh:min].
PLEASE NOTE:
The function is deactivated if dur.t = InF (In no Function) has been set.
You are in the
parameter level
.
rEG
appears on the upper display.
SLor
flashes on the lower display.
Use the
or
key to select the parameter
dur.t
.
rEG
appears on the upper display.
SLoF
flashes on the lower display
Press
the
P
key.
dur.t
flashes on the upper display.
The respective
currently set duration time
ap-
pears on the lower display.
Press
the
key
to
increase
the duration time
dur.t
.
Press
the
key
to
decrease
the duration time
dur.t
.
Press the
and
key to raise and lower the value by 0.1 respectively. If the keys are
held pressed for at least one second, the value increases or decreases quickly and after
two seconds even more quickly; this means the desired value can be reached rapidly.
Press
the
P
key to confirm the set duration time dur.t.
The display returns to the parameter level.
Automatic return!
If no key is pressed for approx. 15 seconds, the device automatically returns to a previ-
ous level up to the
calibration mode
.
Activate temperature profile!
After carrying out the settings, the calibrator / micro calibration bath uses the new values
only when modifying the set temperature or switching the calibrator / micro calibration
bath off and on again.
Parameters for heating gradient SLor
Parameters for the duration time dur.t
Duration time entry