Operating the Calibrator / Micro Calibration Bath
© SIKA • Ba_TP17-TPM-Multi_en.doc • 01/2014
In calibration mode, the upper display will then show the chosen linearisation (LI, DB, Ir or
SU) every 5 seconds in alternation with the actual temperature.
The internal references are set to normal when operating the micro-bath, the dry block and
the infrared function; the selector switch should be turned to int. Ref.
If operated as a surface temperature calibrator then an external reference is used. Please
connect this to the mini-DIN socket and turn the selector switch to ext. Ref
Menu structure S… versions
As displayed by the menu structure, it is possible to reach the
group
and
parameter levels
to carry out settings via
OPEr
.
PLEASE OBSERVE:
Many of the described settings can only be carried out in the S version, but this is dis-
played in the chapter heading.
Returning to another level
If no key is pressed in the
main menu
at the
group
or
parameter level
for approx. 15 sec-
onds, the device automatically returns to the previous level up to the
calibration mode
.
You can also return to a previous level by pressing and holding the
or
key.