Operating the Calibrator / Micro Calibration Bath
7.3.3 Fixed set temperatures (S version)
Setting and saving fixed set temperatures (S version)
In order to save set temperatures in the calibrator / micro calibration bath, the respective set
point memory has to be opened
Press
the
P
key for approx. 5 sec. when in
calibra-
tion mode,
the main menu opens
OPEr
appears on the upper display.
LED SET flashes on the lower display.
Press
the
P
key again, the
group level
opens.
OPEr
appears on the upper display.
SP
appears on the lower display and LED SET
flashes.
Press
the
P
key again, the
parameter level
opens.
SP appears on the upper display.
The set point memory SP 1 and LED SET flash on
the lower display.
Use
the
or
key to select one of the four set
point memories SP1, SP2, SP3 and SP4.
Press
the
P
key to open the respective set point
memory.
The selected set point memory, e.g. SP 3 flashes
on the upper display.
The corresponding current set temperature ap-
pears on the lower display.
Press
the
key to
increase
the set temperature.
Press
the
key
to
decrease
the set temperature.
Press the
and
key to raise and lower the value by 0.1 respectively. If the keys are
held pressed for at least one second, the value increases or decreases quickly and after
two seconds even more quickly; this means the desired value can be reached rapidly.
Press
the
P
key to confirm the set temperature.
The set point memory closes and the display returns to the
parameter level
.
Press and hold the
or
key to return to the
calibration mode
.
If no key is pressed for approx. 15 seconds, the device automatically returns to a previous
level up to the
calibration mode
.
operator menu OPEr
Group SP
Parameter for the set memory SP1
Set point memory SP3 entry