Tank:
The max. filling level in the tank is displayed by the upper edge of the aluminium lining.
The max. filling level is ~0,45 litres.
Tub insert:
The maximum fill level with the tub insert is below the fixture for the sleeve changing tool.
The maximum filling amount is approximately 0.32 litres.
4.4.3 Filling the micro calibration bath
Observe § 4.4.2 "Information on filling amounts"!
When filling, leave enough room for thermal expansion, displacement by sensors and
level rise due to stirring.
Remove the transport cover.
Place the magnet stirrer in the tank.
Place the sensor cage inside.
It protects the magnetic stirrer. It also prevents blocking
and ensures proper stirring
.
Using the sleeve changing tool, fit the tub insert in the block or tank
(only if a tub insert is used).
Insert the test specimen into the sensor cage; in this way the volume of
the sensor to be inspected will be taken into account.
Pour the calibration liquid into the tank or tub insert.
Leave enough space for additional rise in the fill level.
WORK COVER!
For calibration use the included work cover.
It ensures stable positioning of the test specimens in the calibration
bath.
Evaporation of the calibration liquid is minimised by the work cover
and the silicone stoppers.
Screw the work cover onto the tank and insert the sensors through the work cover
into the tank.
*
1
Some steps are unnecessary if the tub insert is already filled.
Tank
Bechereinsatz *