Instruction Manual ScrubberGuard
Servicing
68
14830E/2
WHEN
WHAT
WHERE
WHO
Annually, or
more often
as needed
Clean the venting tubes according to
Section 9.2.4
System
User
Replace the pH sensor according to
Section 9.3.3/ Section 9.3.4
pH sensor
Clean the water tank according to Sec-
tion 9.2.3
System
Clean the parts which come into con-
tact with water according to
Section 9.4.4
AS SG (A)
OG SG (A)
Every 2 years Replace the rubber pads according to
Section 9.2.6
System
User
Replace the instrument pump behind
the tank according to Section 9.2.10
System
Replace the inlet / outlet pump capaci-
tor according to Section 9.2.9
System
Every 3 years Replace the UV LED
OG SG (A)
Service technician
Replace the tension springs on the
ScrubberGuard according to
Section 9.2.5
System
User
Every 5 years Replace the solenoid valves according
to Section 9.2.11
System
User
Replace the check valve according to
drawing: ScrubberG/EXSP-BA
System
User
Replace the seals according to drawing:
ScrubberG/Seal-BA
System
User
Replace the inlet / outlet pump
System
Service technician
Every 10
years
Replace the battery according to Sec-
tion 9.4.6
AS SG (A)
OG SG (A)
User
Table 1: Servicing schedule for ScrubberGuard
9.2
Servicing duties on the system
9.2.1
ScrubberGuard checklist
The following checklist is used for maintaining the measuring mode. If something is found,
consult the corresponding section:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
General visual inspection of the pipes:
▪
Are there indications of leaky screw con-
nections?
▪
Are the pipes damaged?
If any of these questions are
answered in the affirmative, seal or
repair the corresponding area. Re-
place the seals or pipes as needed.