Instruction Manual ScrubberGuard
Servicing
76
14830E/2
9.2.4
Cleaning the deaeration tubes
The following describes cleaning a deaeration tube:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Switch off the measuring mode by pressing
the
Measurement ON/OFF
button.
Drain the water tank by pressing the
Drain
tank
button (Section 7.1).
The pump may only be ope-
rated dry for a maximum of 30
seconds. Danger of overheating!
2.
Remove the deaeration tubes to be cleaned
from the pipe by loosening both screw caps
(arrows).
3.
Rinse out the deaeration tubes with water.
The deaeration hoses are rinsed out against
the flow direction. A syringe can prove
useful here.
4.
Reinstall the cleaned deaeration tubes into
the pipe. In doing so, ensure that the deae-
ration hoses are not kinked.
5.
The system is now ready for operation again.