Servicing
Instruction Manual LabScat 2
12112E/2
41
WORKSTEP
ADDITIONAL INFO / IMAGES
4.
Remove the centering mechanism:
Loosen the two knurled screws (circles) and lift
the entire unit out of the water bath.
5.
Clean the centering mechanism under running
water with a brush and rinse.
This centering mechanism is dishwasher-
proof, but the temperature must not exceed
80 °C.
6.
Remove the floater:
Pull the floater upward and hang it over the
rim of the water bath to the outside.