Instruction Manual LabScat 2
Settings
38
12112E/2
8.5
Setting or changing the access code
You can protect the settings of the instrument against unauthorized manipulations by defin-
ing your own access code.
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Configuration
button.
If the desired menu does not
appear, press the arrow bottom
right.
4.
Press the button to the right of the
Access
code
description text.
5.
Enter the access code and confirm with
OK
.
6.
Press the
Meas
button.
Instrument again in normal opera-
tion.
A forgotten access code can be cleared only by a SIGRIST service engineer.
Enter your personal access code here:
8.6
Backup configured data
These measures can be of use to the service engineers for service purposes.
MANIPULATION
ADDITIONAL INFO / IMAGES
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
System info.
button.
If the desired menu does not
appear, press the arrow bottom
right.
4.
In the
User -> SD
and
Expert -> SD
submenus
press the
Copy
function.
The user and expert data are copied
to the microSD card. After a suc-
cessfully completed procedure,
acknowledge with the
OK
button.
5.
Press the
Meas
button.
Instrument again in measuring op-
eration.