2
8350-1 Temperature Transmitter
2
1
4
3
7
6
5
Output -
System Pwr
Loop -
System Pwr
Loop +
Sensr Gnd
(WHITE)
Sensr IN
(RED)
Sensr V+
(BLACK)
Terminals 3 and 4: Open-Collector Output
• Transistor output, programmable (see CALIBRATE menu) as
High or Low setpoint with adjustable hysteresis.
• May be disabled (Off) if not used.
Terminals 5-7: Digital sensor input
See Wiring Tips for more information.
Terminals 1 and 2: System Loop Power
12-24 VDC ±10% system power and current loop output.
Max. loop impedance:
50
Ω
max. @ 12 V
325
Ω
max. @ 18 V
600
Ω
max. @ 24 V
15
S
D
Internal open-collector
output circuit
Output --
Output +
Isolation
2.1. Panel Installation
1. The panel mount transmitter is designed for installation using a 1/4 DIN Punch. For manual
panel cutout, an adhesive template is provided as an installation guide. The cutout must be
92 mm x 92 mm. Recommended clearance on all sides between instruments is 25 mm/1 in.
2. Slide the instrument into the cutout from the front of the panel.
3. Slide the mounting bracket over back of instrument until quick-clips snap into latches on
side of instrument.
4. To remove, secure instrument temporarily with tape from front or grip from rear of
instrument. DO NOT RELEASE.
Press quick-clips outward and remove.
Output -
Outp
ut +
System Pwr
Loop -
System Pwr
Loop +
2
1
4
3
Sensr Gnd
(SHIELD)
Sensr IN
(RED)
Sensr V+
(BLACK
)
7
6
5
quick-clips
gasket
panel
terminals
mounting
bracket
latch
Panel Mount
Installation Detail
3. Electrical
Connections
Caution:
Failure to fully open terminal jaws before removing wire may permanently damage instrument.
Wiring Procedure
1. Remove 0.5 - 0.625 in. (13-16 mm) of insulation from wire end.
2. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
3. Insert exposed (non-insulated) wire end in terminal hole until it bottoms out.
4. Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.
Wiring Removal Procedure
1. Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
2. When fully open, remove wire from terminal.
Wiring Tips:
• Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
• Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
• Seal cable entry points to prevent moisture damage.
• Only one wire should be inserted into a terminal. Splice double wires outside the terminal.
• The cable length from the sensor to the transmitter must not exceed 400 ft (122 m).
• For best performance, ground the sensor SHIELD wire to a local earth ground at a point near the sensor.
2
1