background image

3

Signet 

2820 Series Conductivity Sensor Instruction Manual

64 mm

 2.5 in./

152 mm

 6.0 in./

13 mm

 0.5 in./

126 mm

5.0 in./

2820-X2
sensor tip

2821-X2
sensor tip

50 mm

2.0 in./

21 mm

0.8 in./

8 mm

0.3 in./

15 ft/4.6 m cable (std.)

6 in./
152 mm

4.2 in./
107 mm

2 in./
50 mm

0.8 in./
21 mm

3/4 in. NPT

15 ft/4.6 m cable (std.)

Reversible
fitting assy.
for submersion
mounting

0.5 in./

13 mm

2821-1
sensor tip

2820-1
sensor tip

0.3 in./
8 mm

152 mm

 6.0 in./

13 mm

 0.5 in./

50 mm

2.0 in./

126 mm

5.0 in./

2820-X1
sensor tip

2821-X1
sensor tip

50 mm

2.0 in./

21 mm

0.8 in./

8 mm

0.3 in./

15 ft/4.6 m cable (std.)

2819-1/2820-1/2821-1 
• 2819-1 cell: 

0.01

   2819-1 range: 

0.010 to 100 µS (10 kΩ to 100 MΩ)

• 2820-1 cell: 

0.1

  2820-1 range: 

1 to 1000 µS

• 2821-1 cell: 

1.0

  2821-1 range: 

10 to 10,000 µS

Temperature compensation:  PT1000
Wetted materials:
 • O-rings: 

EPR

 • Insulator 

material: 

PTFE

  •  Electrodes: 

316 stainless steel

Standard fi tting: 

Polypropylene

 • Max. 

pressure: 

6.9 

bar 

(100 

psi)

 • Max. 

temperature: 

100°C 

(212°F)

Optional fi tting: 

316 stainless steel (1/2 in. NPT)

 • Max. 

pressure: 

13.8 

bar 

(200 

psi)

 • Max. 

temperature: 

120°C 

(248°F)

Quality Standard: 

CE

2819-S1/2820-S1/2821-S1/2819-T1/2820-T1/2821-T1 
• 2819-X1 cell: 

0.01

   2819-X1 range: 

0.010 to 100 µS (10 kΩ to 100 MΩ)

• 2820-X1 cell: 

0.1

  2820-X1 range: 

1 to 1000 µS

• 2821-X1 cell: 

1.0

  2821-X1 range: 

10 to 10,000 µS

Sanitary fi tting size: 

1 and 1

1

/

2

  in.

Temperature compensation:  PT1000
Wetted materials:
 • O-rings: 

EPR

 • Insulator 

material: 

PTFE

  •  Electrodes: 

316 stainless steel or titanium

Sanitary fi tting: 

316 stainless steel or titanium

Max. pressure: 

6.9 bar (100 psi)

Max. temperature: 

120°C (248°F)

Quality Standard: 

CE

2819-S2/2820-S2/2821-S2/2819-T2/2820-T2/2821-T2 
• 2819-X2 cell: 

0.01

   2819-X2 range: 

0.010 to 100 µS (10 kΩ to 100 MΩ)

• 2820-X2 cell: 

0.1

  2820-X2 range: 

1 to 1000 µS

• 2821-X2 cell: 

1.0

  2821-X2 range: 

10 to 10,000 µS

Sanitary fi tting size: 

2 in.

Temperature compensation:  PT1000
Wetted materials:
 • O-rings: 

EPR

 • Insulator 

material: 

PTFE

  •  Electrodes: 

316 stainless steel or titanium

Sanitary fi tting: 

316 stainless steel or titanium

 Max. 

pressure: 

6.9 

bar 

(100 

psi)

 Max. 

temperature: 

120°C 

(248°F)

Quality Standard: 

CE

9.  Specifi cations

Alternate wetted materials and overall lengths are available through special order.  Cable length extensions to 100 ft. (30 m) are available 
through special order.  For resistivity measurements above 10 MΩ and/or below 20°C, maximum cable length is 25 ft. (7.6 m).

2819-1

2819-X2

2819-X1

Summary of Contents for 2820-1

Page 1: ...position 3 4 in NPT Hand tighten only Optional fitting kit Hole up Mark hole position Customer supplied pipe tee reducer 1 2 in NPT Hand tighten only O ring O ring Sealant Sealant Signet 5800CR Use three conductor shielded cable for cable extensions up to 30 m 100 ft max Shield must be maintained through cable splice Disable power to instrument before connecting this sensor RED WHITE BLACK SILVER ...

Page 2: ...onal fitting kit 3 4 in NPT 1 2 in NPT 1 Customer supplied pipe Customer supplied coupling Hand tighten only Standard fitting kit Optional fitting kit O ring Sealant Sealant Optional submersion fitting kit 3 2820 390 1 2 1 1 4 turns past finger tight 3 4 in NPT Customer supplied pipe Customer supplied coupling Sealant Sealant 8 2823 Submersible Installation Fill with 3 to 4 in of sealant Attach 3 ...

Page 3: ... S1 2820 S1 2821 S1 2819 T1 2820 T1 2821 T1 2819 X1 cell 0 01 2819 X1 range 0 010 to 100 µS 10 kΩ to 100 MΩ 2820 X1 cell 0 1 2820 X1 range 1 to 1000 µS 2821 X1 cell 1 0 2821 X1 range 10 to 10 000 µS Sanitary fitting size 1 and 11 2 in Temperature compensation PT1000 Wetted materials O rings EPR Insulator material PTFE Electrodes 316 stainless steel or titanium Sanitary fitting 316 stainless steel or...

Page 4: ... 6 9 bar 100 psi Max temperature 95 C 203 F Quality Standard CE 2823 1 Cell 20 0 Range 200 to 400 000 µS Temperature compensation PT1000 Wetted materials O rings EPR Insulator material PTFE Electrodes 316 stainless steel Fitting 316 stainless steel Max pressure 6 9 bar 100 psi Max temperature 150 C 302 F Quality Standard CE 8 0 in 203 mm 15 ft 4 6 m cable standard 3 4 in NPT Reversible fitting ass...

Reviews: