Refrigerated Centrifuge Sigma 3-16KL
Refrigerated Centrifuge Sigma 3-16KL IVD
6 Using the centrifuge
42
Version 03/2014, Rev. 2.6 of 06/11/2015 • sb
Translation of the original operating manual, part no. 07050
Pos: 153 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 060 B etrie b_Spinc ont rol L +S/Spinco ntr ol L (Ze nt6 )06 0-0 030 -00 30- 0010 Pro gra mm speic her n @ 2 7\m od_ 140 531 930 978 1_68 .docx @ 1 930 25 @ 4 @ 1
6.3.3.1 Saving a program
NOTE
This function is only available if the centrifuge is at a standstill.
• Enter the parameters that are to be included in the program.
• Select the menu item "save Prog" and confirm the selection. The
indication "save" flashes in the activated status.
• Select a random storage location from the program selection list. Free
storage locations are indicated by a flashing display. Any storage
locations that are already occupied will be overwritten during the saving
process.
• Save the program in the desired location.
The program is now saved.
NOTE
The functions "Auto Open", "Buzzer", "ProgLock", and "Cycles" cannot be
entered as part of a program.
Fig. 26: Saving a program
Pos: 154 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 155 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 060 B etrie b_Spinc ont rol L +S/Spinco ntr ol L (Ze nt6 )06 0-0 030 -00 30- 0020 Pro gra mm lade n @ 27\ mod _140 531 931 261 9_6 8.doc x @ 1 930 53 @ 4 @ 1
6.3.3.2 Loading a program
• Select the menu item "run Prog" and confirm the selection. The
indication "run" flashes in the activated status.
• Select the desired program and confirm the selection by pressing the
function knob.
The program is now loaded.
Fig. 27: Loading a program
Pos: 156 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1