Refrigerated Centrifuge Sigma 3-16KL
Refrigerated Centrifuge Sigma 3-16KL IVD
6 Using the centrifuge
38
Version 03/2014, Rev. 2.6 of 06/11/2015 • sb
Translation of the original operating manual, part no. 07050
Pos: 139 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 060 B etrie b_Spinc ont rol L +S/Spinco ntr ol L (Ze nt6 )06 0-0 030 -00 20- 0090 Besc hleuni gun gs- und B remsk urv en @ 27\ mod _14 053 1928 137 2_6 8.d ocx @ 1928 29 @ 4 @ 1
6.3.2.5 Acceleration and deceleration curves
Acceleration
This function is used to select an acceleration curve. The system offers 10
fixed, programmed acceleration curves (curves 0-9).
Brake
This function is used in order to select a curve that decelerates the
centrifuge to a standstill. Deceleration curves are inverted images of the
acceleration curves and are labelled with identical numbers. Deceleration
curve no. 0 represents a brakeless deceleration.
Fig. 19: Preselection of a curve; here: preselection of an acceleration curve
NOTE
Details concerning the acceleration and deceleration curves can be found
in .
Pos: 140 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 141 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 060 B etrie b_Spinc ont rol L +S/Spinco ntr ol L (Ze nt6 )06 0-0 030 -00 20- 0100 Aut omatis che D eckelö ffnu ng @ 27\ mod _14 053 192 838 31_ 68.d ocx @ 192 857 @ 4 @ 1
6.3.2.6 Automatic lid opening function ("AutoOpen")
The automatic lid opening function must be activated so that the lid opens
automatically at the end of the operation.
In order to activate the automatic lid opening function:
• Select the "Auto Open" symbol with the cursor and confirm the
selection. The symbol and the bar under the symbol start to flash.
• Activate the function by turning the function knob. The symbol remains
displayed in a permanent manner and the bar continues to flash.
• Turning the function knob further will deactivate the function. In this
case, the symbol disappears but the bar continues to flash.
• Press the function knob in order to activate the desired setting. The bar
remains visible as long as the cursor is placed over the symbol.
Fig. 20: The automatic lid opening function "Auto Open" is activated
Pos: 142 /01 0 Univ ersal mod ule/Seit enwec hsel @ 0\m od_ 120 211 624 4312 _0. docx @ 10 5 @ @ 1