SPS5000X series service manual
Remote compensation
voltage (single wire)
0.6
V
Rise Time
10% - 90% of rated output voltage, rated resistance load
Rated Load
50
mS
No Load
50
mS
Fall Time
90% - 10% of rated output voltage, rated resistance load
Rated Load
50
mS
No Load
500
mS
Transient response time
1 (Time for recovery to within 0.1% + 10mV of its
rated output against current of 50% ~ 100%.)
mS
C.C Mode
Line Regulation
18
32
45
18
mA
Load Regulation
18
32
45
18
mA
Ripple and Noise
r.m.s
27
54
81
27
mA
Current Setting Accuracy
*0.1%±10
*0.1%±30
*0.1%±40
*0.1%±10
mA
Current programming
resolution
1
mA
Current Readback Accuracy
*0.1%±20
*0.1%±40
*0.1%±50
*0.1%±20
mA
Current Readback resolution
1
mA
Temperature coefficient
200ppm/°C from rated output current following 30-minute warm-up.
ppm/
℃
Protection Function
OVP
Setting Range
8-88
V
Setting Accuracy
± (2% of rated output voltage)
OCP
The maximum output current limit of the front output terminal is 10A.
Setting Range
1.5-16.5
3-33
4.5-49.5
1.5-16.5
A
Setting Accuracy
± (2% of rated output current)
OTP
Over temperature alarm and shut off output.
Low AC Input Protection
shut off output
OPP
The over power limit is approximately 105% of the rated output power.
Rising/Falling Voltage Slew Rate: Only applicable if V-I Mode is set to CV Slew Rate Priority.
0.1~160
V/s
Rising/Falling Current Slew Rate: Only applicable if V-I Mode is set to CC Slew Rate Priority.
0.01~30.00
0.01~60.00
0.01~90.00
0.01~30.00
A/s
Output resistance setting
0~6
0~3
0~2
0~6
Ω
Efficiency
100Vac
>77
%
200Vac
>79
%