SPS5000X series service manual
(Eg: CALibrate:MEAS:CURRent 0,3.662,0.414,0.1934,3.562,3.314)
Format: CALibrate:MEAS:CURRent point2,point3,Im2,Iset2,Im3,Iset3
(Eg: CALibrate:MEAS:CURRent 3.662,18,3.562,3.314,15.976,15.590)
Note: if Im2 >= point2 or Isps2 >= point2, if Im2 > Isps2, then
point2 = Im2 + 0.1, otherwise point2 = Isps2 + 0.1
(13) Send command CALibrate:SAVE CH1,CMEAS to save the
calibration data.
(14) Send command CALibrate:STATe CH1,0 to exit calibration mode.
(15) Current calibration completed.
The parameters corresponding to the different SPS models with the above
codes are shown in the following table
Mode
Rated
current(A)
Iset1
Iset2
Iset3
point1
point2
point3
SPS5161X
7.5
0.1
1.5
7
0
1.5
9
SPS5164X
SPS5165X
SPS5051X
10
0.2
2
9
0
3
18
SPS5081X
15
0.2
3
14
0
3
18
SPS5084X
SPS5085X
SPS5162X
SPS5163X
22.5
0.3
4.5
21
0
4.5
27
SPS5041X
30
0.4
6
28
0
6
36
SPS5044X
SPS5045X
SPS5082X
SPS5083X
45
0.6
9
42
0
9
54
SPS5042X
60
0.8
12
56
0
12
64
SPS5043X
90
1.2
18
84
0
18
108