18
Quick Start
CH1/CH2 Series Mode
Instruction:
In the series mode, CH1 and CH2 are linked internally into one channel controlled
by CH1. The output voltage value is twice compared with that of single channel.
Output ratings
0~60V/ 0~3.2A
Operation steps
:
1.
Press the
to start the Series mode, and the indicator light turns bright
2.
Connect the load to the positive terminal of CH2 and the negative terminal of CH1;
3.
Turn on CH1 and set its current value to 3.2A. In the default setup, the instrument works in
“Coarse”, you can turn to “Fine” by pressing
;
Coarse: 0.1V or 0.1A each step;
Fine: the least precision each step;
4.
Turn on CH1and set the output voltage/current value by revolving the multi-function knob;
5.
Press
button to open the output.
Note:
You can identify the current working state “CC” or “CV” by referring to the indicator light.(red
means CV, green means CC)