Flow Measurement

SITRANS F M

Transmitter TRANSMAG 2 with sensor 911/E

3/106

Siemens FI 01 · 2014

Sensor cables between sensor and transmitter

The signal voltage proportional to the flow and present at the 
electrodes of the EMF is only a few µV to mV. Superimposed on 
this are electrochemical interferences resulting from the contact 
between the electrodes and liquid, and which can be up to sev-
eral Volt. Also frequently superimposed are line frequency inter-
ferences, interferences resulting from vibrations on the pipelines 
or signal cables, as well as strong magnetic fields in the vicinity. 
Sufficient shielding must therefore be provided, as well as fixed 
routing of the signal cables (electrode and magnet current ca-
ble) in the case of remote versions. This also applies to devices 
with integral preamplifier (smartPLUG). The cable length be-
tween the sensor and transmitter must not exceed 100 m 
(328 ft).

Attention must also be paid to the cable routing. Signal cables 
must be routed free of vibration, and protected against strong 
magnetic and stray fields. In case of doubt, the sensor cables 
must be routed in earthed steel conduit.

Operating instructions for SITRANS F M TRANSMAG 2 

This device is shipped with a Quick Start guide and a CD containing fur-
ther SITRANS F literature.

All literature is also available for free at: 

http://www.siemens.com/flowdocumentation

Medium conditions

• Process temperature

-20 ... +150 °C (-4 ... 302 °F) 
depending on the liner

Minimum conductivity of medium

• With SITRANS F M 911/E sensors

 1 µS/cm (0.1 µS/cm depending 

on medium)

Design

Weight of transmitter

4.4 kg (9.7 lb)

Remote version

Transmitter must be connected to 
sensor using shielded cable

Maximum cable length

100 m (328 ft)

Housing

Die-cast aluminum, painted

Displays and keypad

General display

LCD, backlid, two lines with 
16 characters each

Multi-display for

Flow, totalizer, flow velocity

Keypad

4 keys for entering parameters

Power supply 

corresponding to rating plate

• AC  supply

100 ... 250 V AC 

15 %, 

47 ... 63 Hz

• Power consumption 

Approx. 120 ... 630 VA, depend-
ing on sensor

Line fuse

100 ... 230 V AC: T1.6A

Magnet current fuse

F5A/250 V

Selection and Ordering data

Article No.

SITRANS F M  electromagnetic 

transmitter TRANSMAG 2

for alternating field, remote version, 
110 ... 230 V AC

7 M E 5 0 3 4 -

7

A A

7

1 -

7

A A 0

Output/communication

4 ... 20 mA with HART protocol

0

PROFIBUS PA connection

1

4 ... 20 mA with HART protocol,
digital input

2

Operator display and keypad

Without

0

With

1

Cable glands

M20/M16 x 1.5

1

½“ NPT

2

Selection and Ordering data

Order code

Additional information

Please add 

“-Z“

  to Article No. and specify Order 

code(s) and plain text.

Strengthened mounting bracket for wall and pipeline 
installation

A02

Measuring range, specify in plain text:
Y01: 0 to ... m

3

/h

Y01

Pulse significance, specify in plain text:
Y02: 0 to ... pulses/l

Y02

Setting of digital outputs, specify in plain text:
Y03: Setting of digital outputs: ... 

Y03

Measuring-point number (max. 8 characters), specify in 
plain text: Y15: ..........

Y15

Measuring-point description (max. 16 characters), 
specify in plain text: Y16: ..........

Y16

Stainless steel tag plate

Y17

Other post-production requirements (add plain text)

Y99

Description

Article No.

• English

A5E00102775

• German

A5E00192774

• Spanish

A5E00135276

• French

A5E00135275

FI01_2014_en_Kap03.book  Seite 106  Montag, 11. November 2013  4:45 16

© Siemens AG 2013

 

 

icenta Controls Ltd 

Tel: +44 (0)1722 41

    Fax: +44 (0)1722 

    e: [email protected]    www.icenta.co.uk 

Summary of Contents for TRANSMAG 2

Page 1: ...ce status limits flow direction frequency output Mode of operation The flow measuring principle is based on Faraday s law of elec tromagnetic induction according to which the sensor converts the flow into an electrical voltage proportional to the velocity of the flow Function The TRANSMAG 2 is a microprocessor based transmitter with a build in alphanumeric display in several languages The trans mi...

Page 2: ...put with PC cou pling module or HART communi cator Protocol HART version 5 1 Digital output Signal Output Configurable as active or passive signals Active signal 24 V DC 24 mA Ri 170 Passive signal Open collector max 30 V DC 200 mA Output configuration Pulse Pulse significance 5000 pulses s Pulse width 0 1 ms Limit frequency 10000 Hz Limits Limits for flow and quantity flow direction alarm Digital...

Page 3: ... Remote version Transmitter must be connected to sensor using shielded cable Maximum cable length 100 m 328 ft Housing Die cast aluminum painted Displays and keypad General display LCD backlid two lines with 16 characters each Multi display for Flow totalizer flow velocity Keypad 4 keys for entering parameters Power supply corresponding to rating plate AC supply 100 250 V AC 15 47 63 Hz Power cons...

Page 4: ...C01 Cover for sensor cable and gasket Die cast aluminum with corrosion resistant Basic Polyester powder coating min 60 m 7ME5933 0AC02 Cover for mains supply communication Die cast aluminum with corro sion resistant Basic Polyes ter powder coating min 60 m 7ME5933 0AC03 Safety clamp for electronic cover with glass plate 7ME5933 0AC01 7ME5933 0AC06 M20 cable gland set for power and output connec ti...

Page 5: ...4 mA Ri 170 Ω active passive Passive open collector max 30 V DC 200 mA 0 4 20mA Load 600 Ω max 15 V Digital output 1 Passive open collector max 30 DC 100 mA Power supply Transmitter Power supply L N PE conductor Sensor Sensor connection Digital output 2 Relay NC NO max 5W 50 V DC 200 mA or Digital input High 11 30 V DC Low 30 5 V DC green yellow white brown Ref 2 Ref 1 Reference coil Electr 2 EL 2...

Page 6: ...ial Defined via the Order code Protection rings for liners Function To protect the edges of liners from abrasion e g gravel sand etc Used mainly with soft rubber liners and for PTFE liners at high temperatures from 100 to 150 C 212 to 302 F Contact with medium Yes please always check resistance to measured medium Material Stainless steel AISI 316Ti 1 4571 optionally Hastelloy C276 2 4819 Material ...

Page 7: ...inal diameter Nominal pressure Permissible media Category DN inch PN MWP psi 15 25 1 40 580 Gases fluid group 1 and liquids fluid group 1 Article 3 3 200 300 8 12 10 145 Gases fluid group 1 and liquids fluid group 1 II 65 250 2 10 16 232 Gases fluid group 1 and liquids fluid group 1 II 40 100 1 4 40 580 Gases fluid group 1 and liquids fluid group 1 II 350 1000 14 40 10 145 Gases fluid group 1 and ...

Page 8: ...m 4 Tantalum 5 Cable glands terminal box Metric Polyamide terminal box 1 NPT Polyamide terminal box 2 Selection and Ordering data Order Code Additional information Please add Z to Article No and specify Order code s and plain text Two earthing grounding electrodes made of stainless steel AISI 316Ti 1 4571 A02 Two earthing grounding electrodes made of Hastelloy C276 2 4819 A04 Two earthing groundin...

Page 9: ... Selection and Ordering data Article No Order code SITRANS F M electromagnetic flowmeter Earthing rings for flow sensor 911E per unit 7 ME 5 9 0 2 7777 777 Liner Hard rubber soft rubber 1 Novolak 7 PTFE 0 Nominal diameter Mat no 1 4571 316 Ti DN 15 A A DN 25 1 C A DN 40 1 E A DN 50 2 F A DN 65 2 GA DN 80 3 H A DN 100 4 J A DN 125 5 K A DN 150 6 L A DN 200 8 MA DN 250 10 N A DN 300 12 P A DN 350 14...

Page 10: ... 0 17 64 8 5 18 74 11 5 25 35 25 0 55 12 26 57 32 27 59 53 28 61 73 34 74 95 38 83 78 68 149 9 81 178 6 Nominal diameter DN 300 DN 350 DN 400 DN 450 DN 500 DN 600 DN 700 DN 750 DN 800 DN 900 DN 1000 12 14 16 18 20 24 28 30 32 36 40 Build in length L1 Hard rubber version Linatex soft rubber version 500 19 68 550 21 65 600 23 62 650 25 59 650 25 59 780 30 71 910 35 83 1040 40 95 1170 46 06 1300 51 1...

Reviews: