Flow Measurement
SITRANS F M
Transmitter TRANSMAG 2 with sensor 911/E
3/106
Siemens FI 01 · 2014
Sensor cables between sensor and transmitter
The signal voltage proportional to the flow and present at the
electrodes of the EMF is only a few µV to mV. Superimposed on
this are electrochemical interferences resulting from the contact
between the electrodes and liquid, and which can be up to sev-
eral Volt. Also frequently superimposed are line frequency inter-
ferences, interferences resulting from vibrations on the pipelines
or signal cables, as well as strong magnetic fields in the vicinity.
Sufficient shielding must therefore be provided, as well as fixed
routing of the signal cables (electrode and magnet current ca-
ble) in the case of remote versions. This also applies to devices
with integral preamplifier (smartPLUG). The cable length be-
tween the sensor and transmitter must not exceed 100 m
(328 ft).
Attention must also be paid to the cable routing. Signal cables
must be routed free of vibration, and protected against strong
magnetic and stray fields. In case of doubt, the sensor cables
must be routed in earthed steel conduit.
Operating instructions for SITRANS F M TRANSMAG 2
This device is shipped with a Quick Start guide and a CD containing fur-
ther SITRANS F literature.
All literature is also available for free at:
http://www.siemens.com/flowdocumentation
Medium conditions
• Process temperature
-20 ... +150 °C (-4 ... 302 °F)
depending on the liner
Minimum conductivity of medium
• With SITRANS F M 911/E sensors
1 µS/cm (0.1 µS/cm depending
on medium)
Design
Weight of transmitter
4.4 kg (9.7 lb)
Remote version
Transmitter must be connected to
sensor using shielded cable
Maximum cable length
100 m (328 ft)
Housing
Die-cast aluminum, painted
Displays and keypad
General display
LCD, backlid, two lines with
16 characters each
Multi-display for
Flow, totalizer, flow velocity
Keypad
4 keys for entering parameters
Power supply
corresponding to rating plate
• AC supply
100 ... 250 V AC
15 %,
47 ... 63 Hz
• Power consumption
Approx. 120 ... 630 VA, depend-
ing on sensor
Line fuse
100 ... 230 V AC: T1.6A
Magnet current fuse
F5A/250 V
Selection and Ordering data
Article No.
SITRANS F M electromagnetic
transmitter TRANSMAG 2
for alternating field, remote version,
110 ... 230 V AC
7 M E 5 0 3 4 -
7
A A
7
1 -
7
A A 0
Output/communication
4 ... 20 mA with HART protocol
0
PROFIBUS PA connection
1
4 ... 20 mA with HART protocol,
digital input
2
Operator display and keypad
Without
0
With
1
Cable glands
M20/M16 x 1.5
1
½“ NPT
2
Selection and Ordering data
Order code
Additional information
Please add
“-Z“
to Article No. and specify Order
code(s) and plain text.
Strengthened mounting bracket for wall and pipeline
installation
A02
Measuring range, specify in plain text:
Y01: 0 to ... m
3
/h
Y01
Pulse significance, specify in plain text:
Y02: 0 to ... pulses/l
Y02
Setting of digital outputs, specify in plain text:
Y03: Setting of digital outputs: ...
Y03
Measuring-point number (max. 8 characters), specify in
plain text: Y15: ..........
Y15
Measuring-point description (max. 16 characters),
specify in plain text: Y16: ..........
Y16
Stainless steel tag plate
Y17
Other post-production requirements (add plain text)
Y99
Description
Article No.
• English
A5E00102775
• German
A5E00192774
• Spanish
A5E00135276
• French
A5E00135275