Weighing
Functions
Procedures:
o
Check all parameters and change if necessary.
o
Send the DS3 to the scale
(Not possible with the wire bridge on terminal P-PR,
the data is not accepted)
o
Perform the scale adjustment
o
Receive DS3 from the scale
Name
Type
Length
Default
Range of Values / Meaning
Adjustment digits for zero
point
WORD
2
5461
Scale zero point (JD0) [ 0 to: 2
16
]
Other definitions not permitted.
For description see chapter
Adjustment digits for
Adjustment weight 1
WORD
2
60074
Adjustment digits for adjustment weight 1 (JD1) [ 0 to: 2
16
]
Other definitions not permitted.
For description see chapter
Adjustment digits for
adjustment weight 2
WORD
2
0
Adjustment digits for adjustment weight 2 (JD2) [ 0 to: 2
16
]
Other definitions not permitted.
For description see chapter
Adjustment weight 1
INT
2
2000
Adjustment weight 1 (JG1)
For description see chapter
Adjustment weight 2
INT
2
0
Adjustment weight 2 (JG2)
For description see chapter
Characteristic value
range
BYTE
1
2
1: Characteristic value range 1mV/V
2: Characteristic value range 2mV/V
4: Characteristic value range 4mV/V
Other definitions not permitted.
For description see chapter
Measurement time
BYTE 1 0
Bit 0 Measurement time
0 = the measurement time is 20 ms (Sample rate 50 Hz)
1 = the measurement time is 33 1/3 ms (Sample rate 30 Hz)
For description see chapter
Bit 2 to 7 not used
Limit frequency for low
pass filter fg
BYTE
1
4
3: fg = 5Hz
4: fg = 2Hz
5: fg = 1Hz
6: fg = 0.5Hz
7: fg = 0.2Hz
8: fg = 0.1Hz
9: fg = 0.05Hz
Other definitions not permitted.
For description see chapter
Filter depth - Average
value filter
BYTE
1
15
Filter depth of average value filter [0 .. 255]
0 and 1: Average value filter deactivated
>1: Filter depth
For description see chapter
Scale Name
STRING[10] 12 “SIWARE
X MS”
Customer specific scale designation
For description see chapter
Min. weighing range
INT
2
20
Minimum Weight for Weighing Range
For description see chapter
Weighing range
INT
2
2000
Maximum Weight for Weighing Range
For description see chapter
Numeral step
INT
2
1
Numeral step (1, 2, 5, 10, 20)
For description see chapter
Standstill Range
INT
2
10
Standstill range in weight unit
For description see chapter
Standstill Time
TIME
4
1000
Standstill time in ms
For description see chapter