SITRANS LG240 - Operating Instructions
44286-EN-171021
Information:
In this operating instructions manual, the optional instrument features
are described. The respective scope of delivery results from the order
specification.
3.2 Principle of operation
The SITRANS LG240 is a level sensor with polished rod probe for
continuous level or interface measurement, particularly suitable for
applications in the food processing and pharmaceutical industry.
Optionally an autoclaved version with separable housing is available.
High frequency microwave pulses are guided along a steel cable or
a rod. Upon reaching the product surface, the microwave pulses are
reflected. The running time is evaluated by the instrument and output-
ted as level.
d
h
1
Fig. 2: Level measurement
1 Sensor reference plane (seal surface of the process fitting)
d Distance to the level
h Height - Level
High frequency microwave impulses are guided along a steel cable or
rod. Upon reaching the product surface, a part of the microwave im-
pulses is reflected. The other part passes through the upper product
and is reflected by the interface. The running times to the two product
layers are processed by the instrument.
Application area
Functional principle -
level measurement
Functional principle - in-
terface measurement
Summary of Contents for SITRANS LG240
Page 2: ...2 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Page 85: ...Notes 85 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Page 86: ...Notes 86 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...
Page 87: ...Notes 87 PBD 51041038 SITRANS LG240 Operating Instructions 44286 EN 171021 ...