Note
Distance and Custom applications
● When "Operation" is set to "Distance", setting the upper calibration point is not required
(will not appear in the wizard).
● Only when "Operation" is set to "Custom", is the upper range value required (will appear
in the wizard).
Note
Output remains active
While the device is being configured, the output remains active and continues to respond to
changes in the device.
Note
Parameter options
In the following wizard steps, a full list of options for each parameter is shown. However, some
options may not appear on the device, based on the application selected.
Operation
Sets operation mode which defines output and local display.
Setting
Level
LEVEL
Space
SPACE
Distance
DIST
Volume
VOL
Volume flow
VFLOW
Custom
CUSTM
Default
Level
LEVEL
Note
Completing a Custom setup
For a custom application ("Operation" set to "Custom"), at least one pair of input and output
breakpoints (parameters "X-value" and "Y-value") must be defined
after the wizard is complete.
Refer to full operating instructions: In Parameter assignment - Local chapter, see menu Setup >
Custom > Customized characteristic curve.
Commissioning
7.2 Local commissioning
SITRANS Probe LU240 with mA/HART
36
Compact Operating Instructions, 10/2018, A5E42673704-AB