
Modbus communication
A.9 Modbus holding registers tables
SITRANS F Ultrasonic flowmeters FST030 (Modbus)
Operating Instructions, 08/2017, A5E37367739-AC
235
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units regis-
ter)
Value range /
Setting options
Access
level
9148
Unsigned / 4 Alarm items 4
Bit encoded selection
of alarm items that
impact the output. See
(Page 295). Applicable
if Status mode = 1.
0
Read /
write
9150
Unsigned / 4 Alarm items 5
Bit encoded selection
of alarm items that
impact the output. See
(Page 295). Applicable
if Status mode = 1.
0
Read /
write
9156
Unsigned / 4 Alarm items 8
Bit encoded selection
of alarm items that
impact the output. See
(Page 295). Applicable
if Status mode = 1.
0
Read /
write
10156
Unsigned / 4 Alarm items 9
Bit encoded selection
of alarm items that
impact the output. See
(Page 295). Applicable
if Status mode = 1.
0
Read /
write
9154
Unsigned / 4 Alarm items 7
Bit encoded selection
of alarm items that
impact the output. See
(Page 295). Applicable
if Status mode = 1.
0
Read /
write
9152
Unsigned / 4 Alarm items 6
Bit encoded selection
of alarm items that
impact the output.
Mainly simulation
alarms. See Alarm
Applicable if Status
mode = 1.
0
Read /
write
9159
Unsigned / 2 Polarity
Logical polarity of sta-
tus output.
0
•
0: Active high level
•
1: Active low level
Read /
write
9160
Float / 4
On delay
Time to delay the lead-
ing edge of the output.
0.0
[s]
0.0 - 100.0
Read /
write
9162
Float / 4
Off delay
Time to delay the trail-
ing edge of the output.
0.0
[s]
0.0 - 100.0
Read /
write
7155
Unsigned / 2 Forced value
Forced value when
forcing is active.
0
•
0
•
1
Read /
write