SITRANS F C MASSFLO
15
SFIDK.PS.028.Q1.02
4. Technical data
4.2 Signal converter MASS 6000 Ex-d
Measurement of
Mass flow [kg/s], volume flow [l/s], fraction [%],
°
Brix, density [kg/m
3
], temperature [
°
C]
Current output
Classified Ex ia, selectable as active or passive outputs. Default setting is passive mode
Current
0-20 mA or 4-20 mA
Load
<
350 ohm
Time constant
0.1-30 s adjustable
Output characteristics
Active mode
(read note 1)
Passive mode
(max. input from external barrier)
(Terminals: 31-32)
U
o
24 V
U
i
30 V
Io
82 mA
I
i
100 mA
Po
0.5 W
P
i
0.75 W
C
o
125 nF
C
i
52 nF
L
o
2.5 mH
L
i
100
µ
H
Digital output
Frequency
0-10 kHz, 50% duty cycle
Time constant
0.1-30 s adjustable
Passive
6-30 V d.c., max. 110 mA,
1 K
Ω
≤
R
load
≤
10 K
Ω
Output characteristics
Active mode
Passive mode
(max. input from external barrier)
(Terminals: 56-57-58)
Not available
U
i
30 V
I
i
100 mA
P
i
0.75 W
C
i
52 nF
L
i
100
µ
H
Relay
(Terminals: 44-45-46)
Type
Change-over relay
Load
30 V / 100 mA
Functionality
Error level, error number, limit, direction
Output characteristics
U
i
: 30 V, I
i
: 100 mA, P
i
: 0.75 W, C
i
: 0 nF, L
i
: 0 mH
Digital input
(Terminals: 77-78)
11-30 V d.c., Ri = 13.6 K
Ω
Functionallity
Start/hold/continue batch, zero point adjust, reset totalizer 1/2, force output, freeze output
Output characteristics
U
i
: 30 V, I
i
: 3.45 mA, P
i
: 0.10 W, C
i
: 0 nf L
i
: 0 mH
Galvanic isolation
All inputs and outputs are galvanically isolated, isolation voltage 500 volts
Cut-off
Low-flow
0-9.9% of maximum flow
Empty pipe
Detection of empty sensor
Density
0 - 2.9 g/cm
3
Totalizer
Two eight-digit counters for forward, net or reverse flow
Display
Background illumination with alphanumerical text, 3 × 20 characters to indicate flow rate, totalized
values, settings and faults. Time constant as current output
Reverse flow indicated by negative sign
Zero point adjustment
Manual via keypad or remote via digital input
Ambient temperature
Operation:
−
20 to +50
°
C
During storage:
−
40 to +70
°
C (Humidity max. 95%)
Communication
Prepared for client mounted add-on modules certified for Ex-use
HART
Active mode
(read note 1 and 2)
Passive mode
(max. input from external barrier)
(Terminals: 91-92)
U
o
6.88 V
U
i
10 V
Io
330 mA
Io
200 mA
Po
0.57 W
P
i
0.5 W
C
o
20 nF
C
i
0 nF
L
o
100
µ
H
L
i
0
µ
H
PROFIBUS PA
Active mode
Passive mode
(read note 3)
(Terminals: 95-96)
Not available
U
i
17.5 V
I
i
380 mA
P
i
5.32 W
C
i
5 nF
L
i
10
µ
H
MASS 6000 Ex-d
- this page has been updated 2004.01