Servo Screen 390
30
E382 E394E 061 01 02 02
Siemens-Elema AB
4. Service procedures
4
Using a keyboard
A standard AT keyboard can be connected
to the Servo Screen 390.
When the keyboard is connected, it is ready
for use. No setting is needed. The keyboard
can be used as a substitute for the control
knob/pulse generator, but it can also be used
in combination with the control knob/pulse
generator.
The following keys are used to simulate the
control knob:
• Left-arrow key = Turning the control knob
counter-clockwise.
• Right-arrow key = Turning the control knob
clockwise.
• Enter key = Pressing the control knob.
The keyboard can be used when trouble-
shooting a possible malfunction in the
control knob function, for example. If there is
a malfunction when using the control knob/
pulse generator and this malfunction does
not appear when using a connected
keyboard, the error may be in the pulse
generator or in the pulse generator interface.
The keyboard can also be used in combi-
nation with the ”on-screen keyboard” when
entering or changing a patient ID. Only upper
case letters will be displayed in the Patient
ID field.
Summary of Contents for Servo Screen 390
Page 1: ...Servo Screen 390 Service Manual E382 E394E 061 01 02 02 ...
Page 4: ...Servo Screen 390 4 E382 E394E 061 01 02 02 Siemens Elema AB Notes Notes ...
Page 10: ...Servo Screen 390 10 E382 E394E 061 01 02 02 Siemens Elema AB Notes 1 Introduction 1 ...
Page 41: ...Servo Screen 390 E382 E394E 061 01 02 02 41 Siemens Elema AB 6 Index 6 Index ...
Page 44: ...Servo Screen 390 44 E382 E394E 061 01 02 02 Siemens Elema AB Notes 6 Index 6 ...
Page 45: ...Servo Screen 390 E382 E394E 061 01 02 02 45 Siemens Elema AB 7 Diagrams 7 Diagrams ...