MEMOSKOP 2/3/50/100/SUB
RX57-029.061.01
Page 2 of 6
Siemens AG
Rev. 03
10.97
TD SD 34
Medical Engineering
6 - 2
PC board replacement
•
For LITHOSTAR Multiline
- Perform Download according to the instructions RXL2-120.113.01...
- Perform Configuration according to the instructions RXL2-120.032.02...
•
For UROSKOP D1/D2
- Perform Download according to the instructions RLL5-310.113.02...
- Perform Configuration according to page 5-1 of these instructions.
For MEMOSKOP 50
6
Replacing board D2 (memory with SCSI)
6
•
Remove the attachment screws on the two tilt levers of the D2 board.
•
Remove board D2.
•
When removing board D2, be careful not to damage the flat ribbon cable.
•
Tilt the lever down on the flat ribbon cable plug and remove the plug.
•
Connect the plug to the new board.
•
Insert the new D2 board and reattach the tilt levers.
Additional work, depending on the specific system
6
•
On SIREMOBIL 2000
Perform Download according to the instructions RR2-120.061.02...
•
On LITHOSTAR Multiline
Perform Download according to the instructions RXL2-120.113.01...
•
On UROSKOP D1/D2
Perform Download according to the instructions RLL5-310.113.02...
For MEMOSKOP 3
6
Replacing board D21 (memory without SCSI)
6
•
Open the MEMOSKOP.
•
Remove the screws on the two tilt levers of the D21.
•
Remove board D21, insert the new board and reattach the tilt levers.
•
No service work is necessary.
•
Close the MEMOSKOP.